预印本 · 队列研究
重新分类心房颤动消融能量风险特征:基于MAUDE数据库的叙事判定比较分析
Reclassifying Ablation Energy Risk Profiles in Atrial Fibrillation: A Narrative-Adjudicated Comparative Analysis of the MAUDE Database
作者:A. S. Shah, R. Fuller, E. G. Daoud, B. Endo, A. Goodwin, R. Martinez-Parachini, M. Engels, H. I. Lehmann, K. Konstantinidis, R. Ranjan, T. J. Bunch
medRxiv · 2026年10月2日 · Shah 等 11 位作者
不需要生物学背景,多打比方
讲解6 部分
依据论文全文。
这项研究想把心房颤动(一种常见的心律不齐)消融(用一个细管子伸进心脏把乱跳的部位烫掉或烧掉)并发症的真实世界经历,放到过去十年消融方式不断变化的大背景里来看。
“In this study, we sought to place the real-world experience of complications with AF ablation in context of the shifting modalities of performing AF ablation over the past decade”
研究的假设是:脉冲电场消融(用高压电脉冲而不是冷热来消融)的风险面貌与热消融明显不同,而且它这种持续存在的、按器械平台和严重程度分层的模式,在讨论病人风险变化到什么程度可以接受时,值得继续被审视。
“We hypothesized that the PFA risk profile differs markedly from thermal ablation and that its persistent, platform- and tier-specific pattern warrants continued scrutiny when defining an acceptable shift in patient risk”
研究者检索了 MAUDE(美国食品药品监督管理局维护的医疗器械不良事件报告数据库)中2016年1月至2026年6月收到的房颤消融导管报告,把每一份分别归入脉冲电场消融、射频消融或冷冻球囊消融。
“We queried MAUDE for AF ablation catheter reports received January 2016 through June 2026 and assigned each to PFA, radiofrequency ablation (RFA), or cryoballoon ablation”
研究者用一个大语言模型(能读懂和归纳文字的计算机程序)把受伤和死亡报告的文字叙述,判定到预先设定好的并发症类别和严重程度上,并请一个更大的模型复核,还由医生审阅了部分样本。
“We adjudicated injury and death narratives into predefined complication families and severity tiers with a large language model, checked against a larger model and by physician review”
研究者计算了报告比值比(一种比较某类事件在某组中报告得是否更多的指标)和贝叶斯收缩估计(在数据很少时把估计值往中间拉、更稳当的统计方法),并与经验零假设比较,还用中断时间序列(看某个时间点前后趋势是否改变)和前瞻性序贯检测(一边积累数据一边反复检验)做分析,并在2026年的一份保留数据上重复估计。
“We calculated reporting odds ratios (ROR) and Bayesian shrinkage estimators compared against an empirical null, applied interrupted time series and prospective sequential detection, and repeated estimates on a 2026 hold-out”
研究者检索了 FDA 的 MAUDE 数据库,范围是2016年1月至2026年6月收到的报告,按 FDA 收到报告的时间来定日期。
“We queried the FDA MAUDE database for reports received from January 2016 through June 2026, dated by FDA receipt”
研究者检索了五个产品代码,用来覆盖房颤消融所用的导管,分别是 QZI、OAE、LPB、OAD 和 DRA。
“We searched five product codes that encompassed the catheters used for AF ablation (QZI, OAE, LPB, OAD, and DRA)”
作为敏感性分析(换一种算法看结论是否照样成立),研究者把每一类并发症单独拿出来看,重点放在最常见的脑血管、冠状动脉、心包和心律相关事件上。
“As a sensitivity analysis, we examined each complication family on its own, focusing on the most common complications of cerebrovascular, coronary, pericardial, and arrhythmia events”
在44,623份报告中,30.1%是脉冲电场消融、50.9%是射频消融、19.0%是冷冻球囊消融,两个模型对事件是否发生的判断一致率为91.8%(𝜅 = 0.83)。
“Among 44,623 reports, 30.1% were PFA, 50.9% RFA, and 19.0% cryoballoon, with 91.8% model agreement on event occurrence (𝜅 = 0.83)”
脉冲电场消融的报告比值比为:溶血(红细胞被破坏)527(95%置信区间33至8,509)、心律事件2.83(2.55至3.16)、冠状动脉事件2.83(2.30至3.48)、脑血管事件2.39(2.10至2.73);相比之下,食管事件为0.024(0.008至0.075)、心包事件为0.34(0.32至0.37)。
“PFA carried an ROR of 527 (95% CI, 33, 8,509) for hemolysis, 2.83 (2.55, 3.16) for arrhythmia, 2.83 (2.30, 3.48) for coronary, and 2.39 (2.10, 2.73) for cerebrovascular events, against 0.024 (0.008, 0.075) for esophageal and 0.34 (0.32, 0.37) for pericardial events”
脑血管事件报告偏多最明显的是 VARIPULSE 这款器械,但把它去掉后,这一类的比例报告比仍有1.36(1.13至1.65),与 EMBOL-AF 研究中测得的1.56方向一致。
“The cerebrovascular excess was largest for VARIPULSE, but without it the class proportional reporting ratio was 1.36 (1.13, 1.65), concordant with the 1.56 measured in EMBOL-AF”
在这份保留数据中,脑血管、心律和溶血这三个信号得到了重复验证,冠状动脉事件保持了原有方向,但没有达到判定为信号的规则。
“In the hold-out, the cerebrovascular, arrhythmia and hemolysis signals replicated, and coronary events kept their direction without meeting the signal rule”
前瞻性检测在 AVANT GUARD 公布其神经系统事件之前13个月,就标记出了脑血管事件。
“Prospective detection flagged cerebrovascular events 13 months before AVANT GUARD published its neurologic events”
与其他能量来源相比,脉冲电场消融的报告比值比为溶血527(95%置信区间33至8,509)、心律事件2.83(2.55至3.16)、冠状动脉事件2.83(2.30至3.48)、脑血管事件2.39(2.10至2.73)、呼吸事件1.79(1.44至2.24),这些在假发现率校正后都达到了信号判定规则(图2)。
“Against the reporting of other energy sources, PFA carried an ROR of 527 (95% CI, 33, 8,509) for hemolysis, 2.83 (2.55, 3.16) for arrhythmia, 2.83 (2.30, 3.48) for coronary events, 2.39 (2.10, 2.73) for cerebrovascular events, and 1.79 (1.44, 2.24) for respiratory events, each meeting the signal rule after false discovery correction (Figure 2)”
食管损伤0.024(0.008至0.075)、心包事件0.34(0.32至0.37)和死亡0.77(0.67至0.89),在脉冲电场消融中报告得更少。
“Esophageal injury, 0.024 (0.008, 0.075), pericardial events, 0.34 (0.32, 0.37), and death, 0.77 (0.67, 0.89), were reported less often with PFA”
全部70例经确认的溶血事件都来自脉冲电场消融,其中15例仅化验异常、31例出现急性肾损伤但不需要透析、24例需要透析。
“All 70 confirmed hemolysis events were PFA, 15 laboratory only, 31 with acute kidney injury without dialysis, and 24 requiring dialysis”
作者明确表示,他们的研究有几项局限性需要加以考虑。
“Several limitations need to be considered for our study”
一项重要局限是 MAUDE 数据缺少做手术总例数的分母,作者尝试用理论发生率和理论分母来弥补这个不足。
“One important limitation with the interpretation of MAUDE data is the lack of a procedural denominator, which we attempted to resolve by providing theoretical rates and denominators”
作者指出的另一个主要问题,是 MAUDE 和 NCDR(美国国家心血管数据注册)两个数据库都存在事件漏报。
“Another major issue is underreporting of events, both in MAUDE and in the NCDR database”
作者还指出,记录本身既不完整又带着噪声,不同的统计方法对同一类并发症会在不同时间报警、或者根本报不出来(表S2),因此事先无法确定该听哪一个警报。
“The record is also incomplete and noisy, and different statistical approaches flagged the same complication family at different times or not at all (Table S2), leaving open which alarm to act on prospectively”
另外,这么大的记录量靠人工判定会耗时极多,所以作者用了大语言模型;尽管有第二个模型复核和医生抽阅样本,这个模型在全部记录上的准确性仍然不会完美。
“Additionally, manual adjudication of a record this large would be extremely time-intensive, so we used an LLM, whose accuracy across the full record will be imperfect despite a second-model check and physician review of a sample”
作者说明,他们对每一份报告都做了文字叙述层面的判定,并与经验零假设和十年热消融基线作比较,以区分真正的类别效应和只是报告数字上的假象。
“here we adjudicate every report at the narrative level and compare against both an empirical null and a ten-year thermal baseline to separate a genuine class effect from disproportionality artifacts”
作者用一个更大的模型 GPT-5.6 Luna(OpenAI)对分层抽样的报告做对抗式判定,来测量大语言模型判读结果的可靠性。
“We measured the reliability of the LLM findings with a larger model, GPT-5.6 Luna (OpenAI), which performed an adversarial adjudication of a stratified sample of reports”
作者对 Fisher 精确检验的 P 值应用了 Benjamini-Hochberg 方法,以在不同时间段内校正假发现率。
“We applied the Benjamini-Hochberg procedure across the Fisher exact P-values to correct for false discovery rates over different time periods assessed”
本研究使用的是美国 FDA 数据库中已去标识的器械不良事件报告,而非直接招募受试者,因此不涉及对患者的直接干预;但原文没有说明数据使用是否经过伦理审查或豁免,建议核实该数据库研究与所在机构伦理要求的适用关系。AI 分析
研究中的事件判读由大型语言模型完成,虽经第二个模型对抗性复核并由医生复核样本,仍存在误判可能;审查类似方案时需要关注自动化判读的验证流程、误判对结论的影响以及复核记录是否可追溯。AI 分析
原文提到多个模型均来自同一家商业公司(OpenAI)并依赖厂商数据库,属于方法学上的利益关联;建议核实研究者是否声明了与器械厂商的财务关系及是否存在利益冲突。AI 分析
由于 MAUDE 存在漏报且缺少手术例数分母,报告中的比例不代表真实发生率;在向伦理委员会或患者沟通这类风险信号时必须明确这一限制,避免将“报告偏多”误读为“风险必然更高”。AI 分析
研究结论可能影响临床术式选择和患者知情同意内容,建议关注信号发现后如何与监管机构、临床医生沟通,以及患者是否能公平获得不同消融技术及其安全性信息。AI 分析
- 心房颤动(AF)
- 最常见的一种心律不齐,心脏上半部分乱跳,容易形成血栓、引起中风。消融就是设法把乱放电的部位处理掉,让心律恢复规整。
- 脉冲电场消融(PFA)
- 一种较新的消融方式,用高压电脉冲让心肌细胞死亡,而不用冷或热。文章说它理论上对周围组织伤害更小,但报告显示有溶血、冠状动脉痉挛和神经系统事件。
- 热消融(射频消融 RFA / 冷冻球囊消融 CA)
- 传统消融方式:射频消融用加热、冷冻球囊消融用冷冻来破坏乱放电的心肌组织,是脉冲电场消融的对照。
- MAUDE 数据库
- 美国食品药品监督管理局收集的医疗器械不良事件报告库,任何人(厂家、医院、医生)都可以上报,但没有做手术总例数这类分母数据。
- 报告比值比(ROR)
- 一种比较指标:某类事件在脉冲电场消融报告里出现的比例,相对于其他消融方式高多少。大于1说明报告得更多,小于1说明报告得更少。它反映的是报告情况,不等于真实发病率。
- 大语言模型(LLM)判定
- 用能读懂文字的计算机程序通读每一份报告的叙述文字,判断里面记录的是哪一类并发症、有多严重,来代替人工阅读海量报告。
- 前瞻性信号检测
- 一边不断有新报告进来、一边反复做统计检验,看什么时候某项并发症的报告开始异常增多,目的是比论文或注册研究更早发现风险苗头。
- VARIPULSE 与 AVANT GUARD
- VARIPULSE 是其中一款脉冲电场消融导管,其脑血管事件报告最多;AVANT GUARD 是一项临床试验,曾因另一款导管手臂中神经系统事件增多而暂停入组。
动画3 个
1. 研究流程:多少人参加、怎么分组已核对原文
从 FDA MAUDE 数据库查询到 2016 至 2026 年间 44,623 份房颤消融导管报告。
2. 关键结果(数字来自原文)已核对原文
全部报告中,脉冲场消融占 30.1%,射频消融占 50.9%,冷冻球囊消融占 19.0%。
3. 利益相关方:谁承担风险、谁获益已核对原文
患者接受脉冲场消融手术,承担溶血、冠脉痉挛等潜在风险。
Front matter
Reclassifying Ablation Energy Risk Profiles in Atrial Fibrillation A Narrative-Adjudicated Comparative Analysis of the MAUDE Database
Anish S. Shah, MD, MS1,2, Reese Fuller, BS3, Emile G. Daoud, MD4, Bobby Endo, MD, PharmD1, Ashley Goodwin, MD1, Roberto Martinez-Parachini, MD1, Marc C. Engels, MD, PhD1, Immo Lehmann, MD, PhD1, Klitos Konstantinidis, MD1, Ravi Ranjan, MD, PhD1,2, and T. Jared Bunch, MD, FHRS1
1 Department of Medicine, Division of Cardiology, University of Utah, Salt Lake City, UT 2 Department of Medicine, George E. Wahlen VA Medical Center, Salt Lake City, UT
S4 Medical
S4 Medical / PFA Safety Signals in MAUDE
Word Count:
~4500 words Keywords:
Atrial Fibrillation Catheter Ablation Pulsed Field Ablation Product Surveillance Complications Machine Learning Address for correspondence:
T. Jared Bunch, M.D.
University of Utah School of Medicine Division of Cardiology, Department of Internal Medicine 50 N Medical Drive Salt Lake City, Utah 84132 Phone: 801-585-7676
摘要Abstract
Background. Pulsed field ablation (PFA) has been rapidly adopted for atrial fibrillation (AF) on the premise of greater safety than thermal ablation. Real-world signals of hemolysis, coronary spasm, and neurologic events have since emerged, and prior Manufacturer and User Facility Device Experience (MAUDE) analyses relied on term matching without prospective signal detection.
Methods. We queried MAUDE for AF ablation catheter reports received January 2016 through June 2026 and assigned each to PFA, radiofrequency ablation (RFA), or cryoballoon ablation. We adjudicated injury and death narratives into predefined complication families and severity tiers with a large language model, checked against a larger model and by physician review. We calculated reporting odds ratios (ROR) and Bayesian shrinkage estimators compared against an empirical null, applied interrupted time series and prospective sequential detection, and repeated estimates on a 2026 hold-out.
Results. Among 44,623 reports, 30.1% were PFA, 50.9% RFA, and 19.0% cryoballoon, with 91.8% model agreement on event occurrence (𝜅 = 0.83). PFA carried an ROR of 527 (95% CI, 33, 8,509) for hemolysis, 2.83 (2.55, 3.16) for arrhythmia, 2.83 (2.30, 3.48) for coronary, and 2.39 (2.10, 2.73) for cerebrovascular events, against 0.024 (0.008, 0.075) for esophageal and 0.34 (0.32, 0.37) for pericardial events. The cerebrovascular excess was largest for VARIPULSE, but without it the class proportional reporting ratio was 1.36 (1.13, 1.65), concordant with the 1.56 measured in EMBOL-AF. In the hold-out, the cerebrovascular, arrhythmia and hemolysis signals replicated, and coronary events kept their direction without meeting the signal rule. Prospective detection flagged cerebrovascular events 13 months before AVANT GUARD published its neurologic events.
Conclusions. PFA showed disproportionate reporting of hemolysis, cerebrovascular, coronary, and arrhythmic events across sequential testing. Machine learning methods allowed for flagging these signals months before confirmatory trials, demonstrating a scalable tool for early detection of procedural risk.
CLINICAL PERSPECTIVE What is Known Pulsed field ablation has been adopted rapidly for atrial fibrillation on the premise of fewer collateral injuries than thermal ablation.
Hemolysis, coronary spasm, and neurologic events have since emerged in case series, registries, and a paused trial, while prior analyses of Food and Drug Administration device reports relied largely on term matching without narrative adjudication at scale.
What the Study Adds In injury and death reports adjudicated at the narrative level, pulsed field ablation was reported more often than thermal ablation for hemolysis, arrhythmia, coronary, and cerebrovascular events and less often for esophageal and pericardial injury.
The cerebrovascular reporting excess was largest for the VARIPULSE catheter but persisted with that platform removed and concentrated in minor stroke and systemic embolism, suggesting an embolic signal not confined to one platform.
Sequential detection run on reports adjudicated by a large language model would have flagged the cerebrovascular signal 13 months before AVANT GUARD published its neurologic events, suggesting a role for automated surveillance of new ablation technologies.
NON-STANDARD ABBREVIATIONS AND ACRONYMS Abbreviation Definition AF Atrial fibrillation CA Cryoballoon ablation CUSUM Cumulative sum EB05 Empirical Bayes 5th percentile EBGM Empirical Bayes geometric mean FDA Food and Drug Administration IFU Instructions for use LLM Large language model MAUDE Manufacturer and User Facility Device Experience MaxSPRT Maximized sequential probability ratio test NCDR National Cardiovascular Data Registry PFA Pulsed field ablation PRR Proportional reporting ratio RFA Radiofrequency ablation ROR Reporting odds ratio
引言Introduction
Pulsed field ablation (PFA) for cardiac tissue uses selective waveforms to induce irreversible electroporation to ablate conducting myocardial tissue, rather than thermal ablative energy. In theory, this selectivity offers an improved safety profile over thermal ablation technology.1 The first generation of catheters are now broadly available after a rapid sequence of United States approvals, with PulseSelect (Medtronic) in December 2023, FARAPULSE (Boston Scientific) in January 2024, Affera Sphere-9 (Medtronic) in October 2024, VARIPULSE (Biosense Webster) in November 2024, and Volt (Abbott) in December 2025. Large post-approval registries have reproduced the efficacy of pulmonary vein isolation for atrial fibrillation (AF) seen in the original pivotal trials and have reported a safety and efficacy over that of radiofrequency ablation (RFA) and cryoballoon ablation (CA), with no esophageal injury and no persistent phrenic nerve palsy.2–4 MANIFEST-US has since extended that experience with the pentaspline catheter to over 40,000 patients.5
The reported safety profile rests on data that differs from clinical experience in several critical ways. The largest of these registries are associated with the manufacturer, with site selection, event evaluation, and adjudication organized around the catheter under the respective study.2–4 Similarly, the pivotal trials are appropriately sponsored and enroll selected patients with experienced operators, and are generally powered for composite safety and efficacy endpoints.6,7 Composite outcomes can bias signal away or towards the null depending on how endpoints are specified, as seen in non-inferiority trials.8 Because these trials often represent a best case scenario, the real-world experience can unearth complications that are underrepresented and identify the crucial need for prospective signal detection to help respond to risks and patterns that can emerge at scale.
Specific risks and event types have surfaced in smaller series and single-platform evaluations with PFA, the most common being hemolysis with subsequent kidney injury9–13, and coronary spasm following energy application proximal to the coronary arteries.14–16 These risks have been assessed and confirmed prospectively as well in NEMESIS-PFA, which found myocardial injury, hemolysis, and renal dysfunction more prominent than with RFA.17 Neurologic events have instead been more prominent in the regulatory records, the most well known of which is the increased stroke risk seen with VARIPULSE, leading to a pause in January 2025 with subsequent irrigation instructions for use (IFU) corrections that summer.18 Similarly, in the AVANT GUARD trial, there was a pause in enrollment in October 2024 after an early increase in neurologic events in the FARAPULSE arm, subsequently reported upon publication in April 2026.19 The VARIPULSE signal was replicated in a real-world assessment of events in the Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database.18,20 The increased embolic risk has subsequently started to be noted in other registries and single-center studies.21,22
The delay between real-world data signals and publication data leads to a critical gap in safety monitoring, such as when the Sprint Fidelis lead fracture signal was identified a year before its recall.23 The purported solution, such as the MAUDE database, however remains limited by the lack of a public denominator, underreporting, lack of event ascertainment and adjudication, and the idiosyncratic issues that may arise that are platform-specific. The sheer volume of MAUDE exceeds what can be read by hand, term matching of complication codes is insufficient to identify underlying signals outside of smaller, less common procedure types,24 and understanding when and which signals warrant intervention in a noisy baseline are both difficult to perform and interpret. However, prospective signal detection analyses in MAUDE should not be abandoned. In order to identify potential early signs of complications, and allow for more rapid shifts in clinical practice when appropriate, we must have a thorough and thoughtful approach in how to evaluate data in a prospective manner.18
In this study, we sought to place the real-world experience of complications with AF ablation in context of the shifting modalities of performing AF ablation over the past decade. We utilized the MAUDE database to not only demonstrate differences and disproportionality in reporting patterns, but also assess the importance of prospective signal detection methods in ways that are often under-recognized in the literature. Moreover, we sought to contextualize how ablation complications are also responsive to not only energy modality, but the force of external factors, from FDA advisories, preliminary data in clinical trials, and regulatory changes. We hypothesized that the risk profile with PFA is markedly different than with thermal ablation, and that the persistent pattern of complications warrant continued scrutiny when considering how to define acceptable shifting of risk in patient outcomes.
Prior MAUDE evaluations of PFA catalogued adverse events and largely attributed the neurologic signal to a single platform, but each was constrained by term matching without narrative confirmation, by early-adoption reporting windows that predated the current device landscape, and by the absence of a thermal comparator; limitations that risk both misclassifying event types and mistaking notoriety bias for a true class effect. These are not incremental gaps: they directly determine whether the reported neurologic and hemolytic signals are real, generalizable, and attributable to the modality itself, which is precisely the question clinicians and regulators now face and will face with rapidly evolving PFA technologies. A repeat analysis is therefore necessary to resolve them, and here we adjudicate every report at the narrative level and compare against both an empirical null and a ten-year thermal baseline to separate a genuine class effect from disproportionality artifacts. We further use MAUDE to test the under-recognized value of prospective signal detection quantifying how many months sooner routine surveillance could have flagged risks that trials and registries later confirmed, and to examine how complication reporting responds to external forces such as FDA advisories and revised IFU. We hypothesized that the PFA risk profile differs markedly from thermal ablation and that its persistent, platform- and tier-specific pattern warrants continued scrutiny when defining an acceptable shift in patient risk.
方法Methods
We queried the FDA MAUDE database for reports received from January 2016 through June 2026, dated by FDA receipt. Reports from 2015 and prior were excluded as patient problem terms were absent in the preceding years, limiting an equivalent adjudication. We searched five product codes that encompassed the catheters used for AF ablation (QZI, OAE, LPB, OAD, and DRA). Every report carries an FDA event type, which was split into injury or death reports, in which a patient outcome is recorded, and malfunction reports, which describe a device issue, and may not have a patient outcome. Injury or death reports were the basis for all clinical estimands, as malfunction reports do not regularly contain a clinical outcome, and malfunction reports were the basis for all device event estimands. The workflow is shown in Figure 1.
We assigned each report to PFA, RFA, or CA by product code, and compared across all three energy modalities. PFA reports were assigned a platform, restricted to the four with the largest market share:
FARAPULSE (Boston Scientific), PulseSelect (Medtronic), VARIPULSE (Biosense Webster), and Affera Sphere-9 (Medtronic). When the code could not distinguish between the energy source or platform, we manually reviewed the report to identify the brand. MAUDE codes each report with International Medical Device Regulators Forum terms chosen by the reporter at submission, either patient problem terms for injury and death, or device problem terms for malfunctions.
The clinical content of a report is comprised of the patient problem term and the associated narrative text, entered by the manufacturer or by a voluntary reporter such as a facility or a physician. We defined complication families from the definitions used in consensus statements and the large PFA registries,3,25,26 and subdivided each into severity tiers, as seen in Table S1. We then performed event adjudication using both the specified problem terms and the narrative text, with structured instructions and complication definitions. The problem terms were used to select potential candidate complication families, and a large language model (LLM), GPT-5 Nano (OpenAI), read the narrative once for each candidate family and returned a structured answer for every complication tier, with an insufficient information outcome that did not count towards the event count. The report term that was selected at submission was only counted when the narrative confirmed it, which differentiates this analysis from keyword or regular expression matching approaches. We measured the reliability of the LLM findings with a larger model, GPT-5.6 Luna (OpenAI), which performed an adversarial adjudication of a stratified sample of reports. We report the raw agreement and Cohen’s kappa on whether an event occurred, and a physician reviewed the sample.
Our statistical approach was based on additive signal detection strategies. The exposure was energy source and the outcome was a confirmed complication family. Every estimate describes reporting rather than incidence, as MAUDE provides no procedural denominator. We report descriptive counts and proportions by energy source and year. We assessed disproportionality overall, by year, and by quarter from the 2 × 2 table of each energy source against all others, as the reporting odds ratio (ROR) and the proportional reporting ratio (PRR).27 For sparse cells, we performed the Haldane and Anscombe correction by adding 0.5 to every cell, to avoid undefined zero cells.28,29 A signal required a PRR ≥ 2 with 𝜒ଶ ≥ 4 and at least 3 events, or an ROR lower confidence bound above 1. We added two shrinkage estimators built for sparse cells, the information component and the empirical Bayes geometric mean (EBGM).30,31 A shrinkage signal required the empirical Bayes 5th percentile (EB05) to be at least 2. We applied the Benjamini-Hochberg procedure across the Fisher exact P-values to correct for false discovery rates over different time periods assessed.32 At baseline, we assessed the thermal complication associations with the established injury rates (e.g. phrenic nerve injury with CA, or esophageal injury with RFA), and paired those against where PFA was expected to be protective, as the expected null. The estimates over the null pairs gave an empirical null, to assess the reporting gap size between comparators.18
We used interrupted time series as a surrogate for a change in signal, fitting segmented regressions of monthly complication counts, offset for the number of reports, so the modeled quantity was the share of reports carrying the complication, both for the pooled series and for PFA against the concurrent thermal series as a control.33 The change point was set at January 2024, the first full month after FDA approval of PFA, with the predicted slope trend prior as the counterfactual.
We additionally ran a surveillance program in a prospective and retrospective manner. Prospectively, we used a Shewhart chart with Laney’s correction for overdispersion, tracking monthly reporting against a 2022 – 2023 baseline,34 a Poisson cumulative sum (CUSUM) chart, and a maximized sequential probability ratio test (MaxSPRT) to compare simultaneous PFA events with thermal reporting events.35,36 A retrospective scan located a change point for each time series by identifying segmented fits with minimal deviance. All estimates were compared with dates of approvals, safety actions, reporting policy changes, and large clinical studies.
We utilized two methods to assess the impact of a synthetic external denominator to estimate the potential risk. We first inverted the disproportionality test (reverse binomial test) by quarter to find the smallest share that PFA would have to account for from all theoretical AF ablations that would allow the PFA share of complication reports to be unsurprising, under a binomial test at 𝛼 = 0.05. The second assumed no under-reporting of PFA events, and treated PFA reports as the numerator over the published rates of complications. That allowed us to hypothesize a theoretical denominator of procedures that would be required to produce the observed number of reports, segregated by the rates available from recent large studies: MANIFEST-17K,3 MANIFEST-PF,2 EU- PORIA,4 EMBOL-AF,22 and NEMESIS-PFA.17 EMBOL-AF served as the calibration anchor because it measured PFA and thermal events in the same centers over the same years with a denominator.
As a sensitivity analysis, we examined each complication family on its own, focusing on the most common complications of cerebrovascular, coronary, pericardial, and arrhythmia events. We additionally examined complications by individual PFA platform and in a leave-one-platform-out analysis that removed each platform in turn from the class estimate, allowing us to assess the effect of FDA advisories and updates to the IFU on reporting rates as a natural experiment.18 Specifically, we examined the VARIPULSE platform, which had an FDA advisory in January 2025 with subsequent June 2025 updates to the IFU for irrigation flow, and FARAPULSE against the neurological event rates found in the AVANT GUARD trial to assess whether the signal was detectable prior to the study’s publication.19 We additionally re-estimated the disproportionality on reports received from January through June 2026 alone, against estimates from reports received through December 2025, as a later period comparison.
Statistical detail and software are in the Supplemental Material. Analyses performed with R version 4.6.1 (R Core Team, Vienna, Austria), and MAUDE querying and retrieval from openFDA were through the card package.37,38
结果Results
We identified 44,623 reports received from 2016 through 2026, of which 13,440 (30.1%) were PFA, 22,701 (50.9%) RFA, and 8,482 (19.0%) CA (Figure 1, Table 1). Within PFA, FARAPULSE accounted for 9,895 reports, PulseSelect for 1,937, Affera Sphere-9 for 928, and VARIPULSE for 491, with 189 reports on other PFA platforms. PFA reports were 34.9% of the reports received in 2024 and 67.3% of those received in 2025. The injury or death reports had 4,705 PFA, 13,153 RFA, and 4,168 CA reports.
Across 319 paired adjudications, the two models agreed on whether an event had occurred in 91.8% (Cohen’s 𝜅 = 0.83), and a physician reviewed the sample. Across the clinical families, the insufficient information outcome was returned for 44.2% of PFA adjudications and 44.1% of CA against 28.7% of RFA, which were not counted towards the event count.
Results / Disproportionality analysis
Against the reporting of other energy sources, PFA carried an ROR of 527 (95% CI, 33, 8,509) for hemolysis, 2.83 (2.55, 3.16) for arrhythmia, 2.83 (2.30, 3.48) for coronary events, 2.39 (2.10, 2.73) for cerebrovascular events, and 1.79 (1.44, 2.24) for respiratory events, each meeting the signal rule after false discovery correction (Figure 2). Esophageal injury, 0.024 (0.008, 0.075), pericardial events, 0.34 (0.32, 0.37), and death, 0.77 (0.67, 0.89), were reported less often with PFA. The EBGM, shrunk toward the null, met the EB05 threshold for hemolysis (4.99, EB05 = 4.08) and arrhythmia (2.16, EB05 = 2.02), while cerebrovascular events (1.99, EB05 = 1.83) and coronary events (2.23, EB05 = 1.96) sat just below it. When evaluating the quarterly reporting, as seen in Figure 3, 100% of hemolysis, 100% of coronary, and 80% of cerebrovascular events had an ROR exceeding 1, while 100% of esophageal and 90% of pericardial quarters sat below an ROR of 1. In comparison, the expected signal of increased risk from thermal ablation for phrenic nerve injury, pulmonary vein stenosis, esophageal damage, and pericardial events was seen consistently.
Results / Signal detection analysis
The period from 2016 through 2023, described here as the thermal era, serves as a baseline of noise for the frequency and size of series breaks, as seen in Figure 4 (A). Thermal series broke 11 times across the 9 complication families available. The strongest signal was in the RFA esophageal series in January 2018 (deviance drop 39). In the PFA era, from 2024 through 2026 as seen in Figure 4 (B), 7 of the 10 complication families available demonstrated a breakpoint, all between January 2024 and December 2024, with the strongest signal in cerebrovascular in December 2024 (deviance drop 40).
The prospective and retrospective signal detection methods identify different time points within the same complication family. Against concurrent thermal reporting, MaxSPRT crossed its threshold for 7 complication families, notably for cerebrovascular events in March 2025, coronary events in December 2024, and arrhythmia in May 2024. Hemolysis could not be tested, as thermal ablation carried no hemolysis reports. For cerebrovascular events, CUSUM alarmed in December 2024, the same month as detected by the break point method, while with the Shewhart method the threshold was met in May 2025. For coronary events, all detection methods signalled in December 2024. The full findings are seen in Table S2. Against the concurrent thermal series (Figure 5), the PFA reporting proportion at market entry was 5.11 (95% CI, 2.91, 8.97) times the thermal proportion for coronary events, 2.03 (1.24, 3.30) for cerebrovascular events, and 0.71 (0.51, 0.99) for pericardial events.
Results / Denominator analysis
As MAUDE provides no procedural denominator, the following results bound it under stated assumptions rather than measure it. For each complication, we determined the threshold PFA adoption rate, the minimum percentage of all AF ablations performed with PFA, at which its observed share of reports would be proportional, and therefore anticipated, under a binomial test at 𝛼 = 0.05 (Figure 6 (A)). In the quarter beginning April 2026, the required share was 76.1% for cerebrovascular events, 76.1% for arrhythmia, 73.9% for vascular access, 73.5% for hemolysis, and 54.7% for coronary events, against 51.4% for pericardial and 0.0% for esophageal events, and for cerebrovascular events it had risen from 50.0% in the quarter beginning January 2025. The estimates also require the assumption that thermal reporting did not change over the same time frame. As a second bound, we computed the number of PFA procedures required to produce the observed reports at the rates published in recent large studies, if every event reached MAUDE, as demonstrated in Figure 6 (B). Based on the published stroke rates, the number of PFA ablations required during 2025 would range from 38,575 by EU-PORIA to 175,618 by MANIFEST-17K, with 87,600 by EMBOL- AF, the calibration anchor.3,4,22 Over the same time frame, based on the published rates of coronary events, the number required would range from 3,607 by NEMESIS-PFA to 163,333 by MANIFEST-PF.2,17
Results / Sensitivity analyses
We additionally broke down events by complication type and by platform. As shown in Figure 7, the cerebrovascular ROR was highest for VARIPULSE (14.11 (95% CI, 10.93, 18.21)), followed by Affera Sphere- 9 (1.59 (1.12, 2.28)) and FARAPULSE (1.39 (1.12, 1.72)), while PulseSelect (1.28 (0.88, 1.87)) was not. With VARIPULSE removed, the class PRR fell from 2.01 (1.69, 2.39) to 1.36 (1.13, 1.65) and stayed above 1 (Figure 7). After the VARIPULSE IFU change, there was a decrease in the proportion of cerebrovascular event reports, with no change seen to other platforms (Figure S1). A break down by severity tiers is shown in Figure 7. The pooled cerebrovascular series broke in December 2024 and MaxSPRT crossed in March 2025, 13 months before the AVANT GUARD neurologic events were published in April 2026,19 and across that break the FARAPULSE cerebrovascular share rose from 2.9% to 6.1% of its injury or death reports.
Re-estimated on reports received from January through June 2026 alone, against estimates from reports received through December 2025, the cerebrovascular, arrhythmia and hemolysis signals replicated, and coronary events kept their direction at 1.74 (95% CI, 0.84, 3.59) without meeting the signal rule (Figure S2).
讨论Discussion
In this analysis of the 44,623 MAUDE reports across three energy modalities from 2016 through 2026, we found that PFA carried a reporting excess for hemolysis, cerebrovascular, coronary, and arrhythmic events against thermal ablation, and a deficit for esophageal and pericardial events. We found that the excess complication risk also mirrored increases in event rates in active trials and registries, preceding their publications by months to years, and were specific to platforms and complication types. Hemolysis and related kidney injury remain the most disproportionately reported event, across PFA platforms, and continues into 2026 without significant change in reporting frequency. Cerebrovascular events were reported in excess on every platform except PulseSelect, and the class PRR of 2.01 (95% CI, 1.69, 2.39) remained elevated at 1.36 (1.13, 1.65) even after removal of VARIPULSE, concordant in size and direction with the ratio of 1.56 measured in EMBOL-AF.22 Every complication family that later carried a signal demonstrated a break point within the first year of approval, and sequential testing showed that platform-specific and severity-specific signal excess were corroborated by hold-out data from 2026. Together, we demonstrated the utility of a prospective signal detection method over that of any one disproportionality analysis, supported by an automated adjudication system, can reveal signal breaks by modality and platform and allow for validation against different historical eras and time points.
With tens of thousands of reports each year, MAUDE can no longer be adjudicated by hand. While some have managed to manually adjudicate small subsets of MAUDE event narratives, most studies are based off of term matching and classification.20,24,39,40 LLM-guided adjudication carries a risk of systematic error, but it offers a powerful means of classifying large datasets with greater speed and consistency than manual review. There are inconsistencies inherent to the MAUDE database, including differences between manufacturers, voluntary and mandatory reporting styles, and the details of the clinical event itself. For example, although tamponade was reported less often with PFA, our analysis detected a signal for an increased reporting rate of pericarditis, two events that might otherwise have been conflated under a single “pericardial injury” term (Figure S3).3,41,42 Within arrhythmia events, additional need for cardioversion, drug therapy, or repeat ablation were noted in the PFA group, while in the RFA group, permanent device implantation was more common (Figure S3). On the other hand, coronary events were confined to only transient spasm, with no excess in sustained spasm, infarction, or stenosis, consistent with the described mechanism.14–16 We were also able to classify report types by platform in addition to event severity. For example, the cerebrovascular event signal of PFA was previously attributed to VARIPULSE alone when compared against other PFA platforms.20,40 In our review, although VARIPULSE had an increased ROR of 14.11 (95% CI, 10.93, 18.21) for cerebrovascular events, the other platforms, with the exception of PulseSelect, still demonstrated increased risk over thermal ablation (Figure 7). The revised VARIPULSE protocol reports fewer complications,43 which we also note after the change in the IFU in June 2025 [Figure S1], however compared to the thermal counterparts, the reporting risk remained elevated, although confounded by notoriety bias. Similarly, the AVANT GUARD trial noted an early increase in stroke rates with FARAPULSE, which was identifiable in the MAUDE database significantly prior to publication of the trial results - that type of safety signal could potentially be used prospectively to modify practice in response, such as increasing anticoagulation levels as part of the procedural protocol. These signals can be slower to surface in industry-sponsored Investigational Device Exemption trials, as these are usually designed with composite safety endpoints rather than specific individual safety endpoints.8
Several limitations need to be considered for our study. One important limitation with the interpretation of MAUDE data is the lack of a procedural denominator, which we attempted to resolve by providing theoretical rates and denominators. The rationale behind this was to assess whether the introduction of PFA was associated with increased adverse event profiles in the real-world experience, expressing reporting excesses as the PFA usage share or procedural volume it would imply, rather than to quantify specific risk rates. By inverting the binomial disproportionality test, we were able to provide an evaluation of the minimum share of total ablations using PFA that would be required to statistically explain the volume of MAUDE reports attributed to PFA (Figure 6). This circumvents the need for estimates of the distribution of procedures across energy sources, which are not reliably known. For example, by the quarter beginning April 2026, at least 76.1% of ablations would have had to be with PFA to explain the excess share of cerebrovascular events over that of thermal. Registry data place PFA adoption at roughly 60% in 2025 among centers reporting to the National Cardiovascular Data Registry (NCDR),44 and those centers are among the earliest adopters, so national adoption is unlikely to have exceeded that. Procedure mix alone therefore does not account for the excess. Another major issue is underreporting of events, both in MAUDE and in the NCDR database. We believe that underreporting bias likely strengthens our findings, as the events captured in the MAUDE database are likely an underrepresentation of the real world experience. To bound the procedure volume these reports imply, we applied the reported complication rates in large studies to the number of PFA event reports, assuming 100% event reporting, to estimate the PFA procedure volume that would have to have been performed for that event rate to hold true. Where the implied volume exceeds the number of procedures plausibly performed, the published literature may be underestimating complication rates, and where it falls short, the gap is more likely evidence of underreporting to MAUDE. This approach allows us to bound the denominator for emerging technologies before procedure-level data are available.
The record is also incomplete and noisy, and different statistical approaches flagged the same complication family at different times or not at all (Table S2), leaving open which alarm to act on prospectively. Additionally, manual adjudication of a record this large would be extremely time-intensive, so we used an LLM, whose accuracy across the full record will be imperfect despite a second-model check and physician review of a sample. These limitations guide the next steps, from improved adjudication methods to prospective implementation of surveillance in MAUDE.
结论Conclusion
In this MAUDE analysis, PFA showed disproportionate reporting of hemolysis, cerebrovascular, coronary, and arrhythmic events, and the cerebrovascular, arrhythmia and hemolysis excesses replicated in reports received from January through June 2026. Large language model adjudication of report narratives made prospective signal detection feasible at scale, surfacing the cerebrovascular signal months before AVANT GUARD published it and suggesting a scalable tool for early detection of procedural risk as new ablation technologies enter practice. We believe that combining this data with clinical trials and registry data will allow us to minimize the exposure of risk to patients.
Tables
Table 1: Composition of the adjudicated report set by year. Counts are reports, not patients or procedures, so no row is an incidence. Percentages are of that year’s column. The complication rows are confirmed adjudications and are not mutually exclusive, since one report can carry more than one family.
Figures
Figure 1: Report selection, machine adjudication, and analysis. Counts are reports, not patients or procedures. The Injury or Death stream is the primary reference set for every clinical estimate; malfunction reports carry a patient outcome too rarely to serve as one and are analysed only for device problems. Adjudication was performed by a structured-output model over every narrative, re-read on the same prompt and gating by a larger model, with a physician-reviewed subsample.
Figure 2: Complication reporting mix, PFA versus thermal ablation. PFA was reported in excess for hemolysis, coronary, arrhythmia and cerebrovascular events and less often for esophageal and pericardial events. (A) Confirmed complications as a share of each energy source’s own injury and death reports, a comparison of reporting mix rather than of complication rates per procedure. (B) Reporting odds ratio for each complication family, PFA versus all other energy sources, with 95% confidence intervals on a log scale; filled points met the prespecified signal rule. The hemolysis and esophageal intervals run past the axis and are pinned at the panel edge rather than dropped. (C) The same contrast by filing stream, where mandatory is by manufacturer and voluntary is by facility or physician; a family with no reports on one arm carries only the other arm’s estimate. All three panels carry the same twelve families in the same order, set by the pulsed field reporting odds ratio. AKI, acute kidney injury; CA, cryoballoon ablation; PFA, pulsed field ablation; PV, pulmonary vein; RFA, radiofrequency ablation.
Figure 3: PFA reporting odds ratios by complication and quarter. Quarters start with the introduction of PFA, and dots represent false discovery rate (FDR) corrected significance. PFA, pulsed field ablation.
Figure 4: Reporting volume and signal timing in the thermal and pulsed field eras. Thermal-era series breaks set the background scale of signal noise for the pulsed field era. (A) Injury and death reports per month for radiofrequency and cryoballoon ablation from 2016 to the first pulsed field approval, and beneath it the month each complication family’s series broke under a retrospective change-point scan, with the interval around each break; a break is a change in slope, not a safety event. (B) Injury and death reports per month by energy source from the first pulsed field approval, and beneath it the month each family first broke (circle, the same scan) or was first declared (diamond, maximum sequential probability ratio test against the concurrent thermal stream). Dated approvals, safety actions and publications are marked on both panels. AKI, acute kidney injury; CA, cryoballoon ablation; IFU, instructions for use; MaxSPRT, maximized sequential probability ratio test; PFA, pulsed field ablation; RFA, radiofrequency ablation.
Figure 5: Complication reporting at PFA market entry against a concurrent thermal control. Pulsed field reporting of hemolysis, coronary, cerebrovascular and arrhythmia events increased at market entry, while pericardial and esophageal reporting decreased. (A) Monthly share of each energy source’s injury and death reports carrying each complication over the last five years, with the segmented fit for each series broken at the first pulsed field approval (dashed). Each strip carries the pulsed field level ratio against radiofrequency and against cryoballoon separately, because cryoballoon carries its own embolic profile and pooling it with radiofrequency moves the comparator. (B) Segmented quasi-Poisson fit of the pooled monthly reporting proportion for the four families of interest, with the pre-period trend carried forward as the counterfactual (dashed), the pinned break at first approval (dashed vertical) and the independently detected break with its interval (dotted, shaded). Colours are the complication palette in (B) and the modality palette in (A), and the two share no chip. CA, cryoballoon ablation; PFA, pulsed field ablation; RFA, radiofrequency ablation.
Figure 6: Procedure volume bounds on PFA complication reports. Procedure volume alone cannot explain the pulsed field share of cerebrovascular, hemolysis, coronary and arrhythmia reports, and under EMBOL-AF, the calibration anchor, the implied procedure volumes fall below plausible real-world volumes, indicating that MAUDE captures a minority of events. (A) The smallest PFA share of procedures in each quarter at which the observed PFA share of reports would no longer exceed expectation under a binomial test at 𝛼 = 0.05; the last quarter is labelled, and quarters with fewer than five reports are omitted. (B) Pulsed field procedures implied in each quarter if every event reached MAUDE, computed as confirmed reports divided by each study’s published rate, with panels ordered by cohort size on a free y-axis. Rates from EMBOL-AF,22 MANIFEST-17K,3 MANIFEST-PF,2 EU-PORIA,4 and NEMESIS-PFA.17 AKI, acute kidney injury; MAUDE, Manufacturer and User Facility Device Experience; PFA, pulsed field ablation.
Figure 7: Cerebrovascular reporting by PFA platform and severity tier. The excess is largest for VARIPULSE but is also seen with other PFA platforms. With VARIPULSE removed the class estimate falls but stays above 1, bracketing the ratio EMBOL-AF measured, with the excess seen in the minor-stroke and systemic-embolism tiers. (A) Cerebrovascular reporting odds ratio for each pulsed field platform against pooled thermal ablation, with the platform’s confirmed events over its injury and death reports. (B) The proportional reporting ratio for the class with every platform and with VARIPULSE removed, and the reporting odds ratio for VARIPULSE alone, beside the pulsed field to thermal incidence ratio measured in the EMBOL-AF registry (vertical line). (C) Pulsed field against pooled thermal within each adjudicated cerebrovascular severity tier; filled points met the signal rule. (D) The same severity contrast within each platform, harder to interpret due to decreased numerators, and classification biased by manufacturer reporting style. CI, confidence interval; PFA, pulsed field ablation; TIA, transient ischemic attack.
SUPPLEMENTAL MATERIAL Supplemental Methods: data source, device problems, event adjudication, disproportionality, and interrupted time series and signal detection Supplemental Results: report characteristics, complication mix by energy source, device problems, severity and platform-specific analysis, and calibration and sensitivity analyses Table S1: Complication families and severity tiers Table S2: Earliest month each signal detection method fired Figure S1: VARIPULSE cerebrovascular reporting before and after the irrigation instructions for use were revised Figure S2: Hold-out re-estimation on reports from January to June 2026 Figure S3: PFA reporting by severity tier within complication families Figure S4: Device problems in malfunction reports Statistical Software: R version, platform, and package versions Supplemental Methods
Figures / Data source
We searched five product codes that encompassed the catheters used for AF ablation: QZI (percutaneous cardiac ablation catheter for atrial fibrillation with irreversible electroporation), OAE (percutaneous cardiac ablation catheter for atrial fibrillation), LPB (cardiac ablation percutaneous catheter), OAD (percutaneous cardiac ablation catheter for atrial flutter), and DRA (steerable catheter). All reports were carried as a sensitivity denominator to assess overall reporting rates.
Figures / Device problems
We took the International Medical Device Regulators Forum (IMDRF) Annex A device problem terms coded on each malfunction report, grouped them by their Annex A family term, and compared each energy source’s problem mentions before and after PFA approval.
Figures / Event adjudication
We paired each family with the IMDRF patient problem terms by matching the full term list through manual physician review. The LLM was given the same prompt and instructions, and defaulted to absent unless the narrative supported the tier, with an insufficient information outcome retained as an alternative. The adversarial sample was stratified by complication type, and GPT-5.6 Luna was given the same definitions and tiers. A physician reviewed the sample for the event labels and complication tiers.
Figures / Disproportionality
The information component is a Bayesian logଶ of observed to expected reports, and the EBGM shrinks the relative reporting ratio towards the null when a cell carries little data.30,31
Figures / Interrupted time series and signal detection
Quasi-Poisson models were fitted on monthly counts to maintain point estimates while accounting for highly dispersed data. Diagnostics were performed on each interrupted time series fit, including lag autocorrelation and a Ljung-Box test of residuals, F test on seasonality, and negative binomial refit to assess variance assumptions. Complications with too few events for signal detection were reported as not estimable.
Figures / Report characteristics
The malfunction reports were 65.0% of PFA reports against 42.0% of RFA and 50.9% of CA. The median time from event to FDA report was 20 to 28 days, and 47.1% of PFA reports originated in the United States against 39.7% of RFA and 31.8% of CA. Manufacturers, as mandatory reporters, filed 96.4% to 98.9% of each modality’s reports, and voluntary reports held 243 PFA reports against 197 thermal reports (Figure 2 (C)).
Figures / Complication mix by energy source
Within each modality, as seen in Figure 2 (A), cerebrovascular events were 8.2% of PFA reports against 3.6% of RFA and 3.7% of CA, arrhythmia 13.4% against 5.3% and 5.0%, and coronary events 3.4% against 1.2% and 1.3%, with hemolysis reported by PFA alone. Pericardial events were 24% of PFA reports against 54% of RFA and 31% of CA, and esophageal injury 0.06% against 3.0% and 1.4%.
Figures / Device problems
In the malfunction reports, PFA had 7,630 IMDRF device problem reports compared to the 4,497 for RFA (Figure S4). The most common device related problem families were material integrity issues (25.7%), followed by device positioning/separating problems (19.5%). The most common specific problems were entrapment of the device itself or material deformation.
Figures / Severity and platform-specific analysis
Within families (Figure S3), the cerebrovascular excess sat in minor non-disabling stroke, 3.39 (95% CI, 2.50, 4.60) on 191 PFA against 52 thermal events, and systemic embolism, 2.21 (1.27, 3.82) on 43 against 18, with transient ischemic attack (TIA) 1.17 (0.83, 1.65), major disabling stroke 1.29 (0.91, 1.83), and fatal stroke 0.80 (0.38, 1.68) null; a tier could not be assigned in 501 PFA and 306 thermal reports, 1.51 (1.32, 1.73), as imaging and follow-up were often absent from the narrative. The coronary excess was confined to transient spasm, 10.47 (5.30, 20.66) on 102 against 9, with sustained spasm 2.77 (0.89, 8.58), myocardial infarction 0.99 (0.65, 1.50), and coronary stenosis 0.35 (0.09, 1.30) null. Every mechanical pericardial tier ran below 1, tamponade treated by pericardiocentesis at 0.48 (0.44, 0.51) and by surgery at 0.56 (0.48, 0.65), while pericarditis without effusion ran the other way, 2.22 (1.36, 3.65) on 53 against 22. Within arrhythmia, selfterminating events 1.91 (1.41, 2.57), events requiring cardioversion or drugs 2.78 (2.25, 3.43), and repeat ablation 1.87 (1.32, 2.64) were reported more often with PFA, and a permanent pacemaker or defibrillator less often, 0.49 (0.39, 0.63) on 94 against 176. All 70 confirmed hemolysis events were PFA, 15 laboratory only, 31 with acute kidney injury without dialysis, and 24 requiring dialysis.
Figures / Calibration and sensitivity analyses
The reference set of expected associations recovered 6 of 6 expected thermal excesses, 1 of 3 expected PFA deficits, and 3 of 6 expected nulls. The 3 null pairs in which no other modality was expected to signal had a geometric mean ROR of 1.03 (SD of the log ROR, 0.23), against 0.68 (1.04) across all 6 expected nulls, too few pairs to fit a calibration distribution, so the estimates were reported uncalibrated. The information component lower bound exceeded 0 for cerebrovascular, arrhythmia, coronary, hemolysis and respiratory events, including cerebrovascular events at 0.84 (0.69, 0.98). With all reports as the denominator rather than injury or death reports alone, the PFA ROR was lower by a factor of 1.35 to 1.70 across families, signal status changed only for respiratory events, and MaxSPRT crossed for cerebrovascular events in February 2026 rather than March 2025. Across the pooled interrupted time series fits, Ljung-Box tests indicated residual autocorrelation for arrhythmia, coronary and hemolysis events, the F test indicated seasonality for infection events, and a negative binomial refit moved no level estimate by more than 12%.
Supplemental Tables Supplemental Figures
Table S1: Complication families and severity tiers as given to the adjudication model. Definitions are reproduced from the schema the model was prompted with, so the table cannot drift from what was actually adjudicated. insufficient_info is an adjudication outcome rather than a severity and is never ordered with the tiers above it.
Table S2: Earliest month that the signal detection methods fired for each complication family. – means that method never was triggered (no signal detected). Shewhart, CUSUM, and MaxSPRT are all prospective, and fire when hitting a specific rule. Series break is retrospective, by looking over past window to find a bestfitting timepoint.
Figure S1: VARIPULSE cerebrovascular reporting before and after the irrigation instructions for use were revised. The VARIPULSE share fell after the revision but remained elevated compared to other platforms. Cerebrovascular share of injury and death reports for each platform and for thermal ablation before and after the VARIPULSE irrigation instructions for use were rewritten on 26 June 2025, with counts at each end. IFU, instructions for use; PFA, pulsed field ablation.
Figure S2: Hold-out re-estimation on reports from January to June 2026. Reporting odds ratio for each complication family estimated on reports received from 2016 through 2025 (circle) and re-estimated on January to June 2026 alone (arrowhead), a period that entered the overall estimates but that no modelling decision was tuned to; colour gives the verdict. AKI, acute kidney injury; PFA, pulsed field ablation; PV, pulmonary vein.
Figure S3: PFA reporting by severity tier within complication families. The coronary excess was confined to transient spasm, every mechanical pericardial tier ran below 1 while pericarditis without effusion ran above it, and the arrhythmia excess did not extend to permanent device implantation. Pulsed field against pooled thermal ablation within each adjudicated severity tier for (A) coronary, (B) cerebrovascular, (C) pericardial and (D) arrhythmia events, with proportional reporting ratios and 95% confidence intervals on one shared log axis so the four panels can be compared by position; filled points met the signal rule, and the estimate with the pulsed field and thermal counts sits at the right of each row. CI, confidence interval; ICD, implantable cardioverter-defibrillator; PFA, pulsed field ablation; TIA, transient ischemic attack.
Figure S4: Device problems in malfunction reports. (A) Device-problem families from manufacturer-coded IMDRF Annex A from MAUDE. No overlap with injury/death reports. (B) Most common PFA device issues broken to individual Annex A terms. IMDRF, International Medical Device Regulators Forum; MAUDE, Manufacturer and User Facility Device Experience; PFA, pulsed field ablation.
Figures / Statistical Software
Analyses were performed with R version 4.6.1 (R Core Team, Vienna, Austria) on macOS Golden Gate 27.0,37 and the manuscript was rendered with Quarto 1.10.18. MAUDE querying and retrieval from openFDA were through the card package.38 The R packages called by the analysis pipeline are listed below with their versions.