非人灵长类硅基皮层内微电极阵列的失效模式分析

J Neural Eng · 2013年11月12日 · Barrese 等 7 位作者

问这篇
一分钟了解要点回顾78枚猴脑植入阵列,多数一年内失效,连接器与绝缘材料是主因。

不需要生物学背景,多打比方

正在获取全文并生成讲解(拿不到全文就依据摘要)…

已等待 0 秒大约需要 10–20 秒

可以先看别的,做好了会自动出现在这里。

这篇还没有动画

动画会把研究的流程、作用机制和关键结果一步一步演示出来,每一步都标明出自原文哪里。制作大约需要一两分钟。

目前只拿到了摘要全文暂时拿不到(可能不是免费全文)。下面是论文摘要。

摘要Abstract

摘要第 1 段问这一段

OBJECTIVE: Brain-computer interfaces (BCIs) using chronically implanted intracortical microelectrode arrays (MEAs) have the potential to restore lost function to people with disabilities if they work reliably for years. Current sensors fail to provide reliably useful signals over extended periods of time for reasons that are not clear. This study reports a comprehensive retrospective analysis from a large set of implants of a single type of intracortical MEA in a single species, with a common set of measures in order to evaluate failure modes.

摘要第 2 段问这一段

APPROACH: Since 1996, 78 silicon MEAs were implanted in 27 monkeys (Macaca mulatta). We used two approaches to find reasons for sensor failure. First, we classified the time course leading up to complete recording failure as acute (abrupt) or chronic (progressive). Second, we evaluated the quality of electrode recordings over time based on signal features and electrode impedance. Failure modes were divided into four categories: biological, material, mechanical, and unknown.

摘要第 3 段问这一段

MAIN RESULTS: Recording duration ranged from 0 to 2104 days (5.75 years), with a mean of 387 days and a median of 182 days (n = 78). Sixty-two arrays failed completely with a mean time to failure of 332 days (median = 133 days) while nine array experiments were electively terminated for experimental reasons (mean = 486 days). Seven remained active at the close of this study (mean = 753 days). Most failures (56%) occurred within a year of implantation, with acute mechanical failures the most common class (48%), largely because of connector issues (83%). Among grossly observable biological failures (24%), a progressive meningeal reaction that separated the array from the parenchyma was most prevalent (14.5%). In the absence of acute interruptions, electrode recordings showed a slow progressive decline in spike amplitude, noise amplitude, and number of viable channels that predicts complete signal loss by about eight years. Impedance measurements showed systematic early increases, which did not appear to affect recording quality, followed by a slow decline over years. The combination of slowly falling impedance and signal quality in these arrays indicates that insulating material failure is the most significant factor.

摘要第 4 段问这一段

SIGNIFICANCE: This is the first long-term failure mode analysis of an emerging BCI technology in a large series of non-human primates. The classification system introduced here may be used to standardize how neuroprosthetic failure modes are evaluated. The results demonstrate the potential for these arrays to record for many years, but achieving reliable sensors will require replacing connectors with implantable wireless systems, controlling the meningeal reaction, and improving insulation materials. These results will focus future research in order to create clinical neuroprosthetic sensors, as well as valuable research tools, that are able to safely provide reliable neural signals for over a decade.