论文 · 早期(1 期)试验
用脑控肌肉电刺激恢复四肢瘫痪者的伸手和抓握动作:概念验证
Restoration of reaching and grasping movements through brain-controlled muscle stimulation in a person with tetraplegia: a proof-of-concept demonstration
作者:A Bolu Ajiboye, Francis R Willett, Daniel R Young, William D Memberg, Brian A Murphy, Jonathan P Miller, Benjamin L Walter, Jennifer A Sweet, Harry A Hoyen, Michael W Keith, P Hunter Peckham, John D Simeral 等 15 人
Lancet · 2017年3月28日 · Ajiboye 等 15 位作者
不需要生物学背景,多打比方
讲解6 部分
依据论文全文。
研究展示了一位因高位颈段脊髓损伤而四肢瘫痪的人,用自己的瘫痪手臂和手,借助功能性电刺激(FES)重新产生动作,并通过植入大脑的脑机接口(iBCI)用他自己的大脑信号来指挥,完成协调的伸手和抓握动作。
“We demonstrate an individual with traumatic high cervical SCI performing coordinated reaching and grasping movements using his own paralyzed arm and hand, reanimated through FES, and commanded using his own cortical signals through an intracortical brain-computer-interface (iBCI)”
研究评估了这位受试者能否用大脑信号指挥他瘫痪的手臂完成简单的单关节手臂/手部动作,以及有实际意义的多关节动作。
“We assessed the participant’s ability to cortically command his paralyzed arm to perform simple single-joint arm/hand movements and functionally meaningful multi-joint movements”
研究首次展示一位慢性四肢瘫痪者使用植入式FES系统,由脑机接口直观且有效地指挥,完成伸手和抓握两种动作,并指出未来走向临床应用的转化路径。
“We for the first time show an individual with chronic tetraplegia using an implanted FES system to make both reaching and grasping movements, intuitively and effectively commanded by an iBCI, with a translational path for future clinical viability”
植入时,他是一位53岁男性,在入组前八年遭遇了创伤性高位颈段脊髓损伤(C4,ASIA A)。
“At the time of implant, he was a 53-year-old man who experienced traumatic high cervical SCI (C4, ASIA A) eight years before enrollment”
这位53岁、C4、ASIA A的受试者在运动皮层手部区域植入了两个皮层内微电极阵列,并植入了36根经皮电极,用于电刺激手、肘和肩部肌肉。
“The study participant (53 years old, C4, ASIA A) received two intracortical microelectrode arrays in the hand area of motor cortex, and 36 percutaneous electrodes for electrically stimulating hand, elbow, and shoulder muscles”
在用脑机接口指挥3D虚拟手臂四个月后,受试者分两次手术(植入后第125天和第280天)在右上下臂植入了36根经皮肌肉刺激电极(包括四根经皮阳极电流回路电极),用来恢复手指和拇指(实现侧向抓握)、腕、肘和肩部动作。
“After four months of using the iBCI to command movements of a 3D virtual arm, the participant received, during two procedures (125 and 280 days-post-implant), 36 percutaneous muscle stimulating electrodes (Synapse Biomedical, Oberlin, OH) in his right upper and lower arm, including four percutaneous anodic current return electrodes, to restore finger and thumb (for a lateral hand grasp), wrist, elbow, and shoulder movements (Supplemental Table 3 lists implanted muscles)”
研究把用脑机接口控制他瘫痪手臂的表现,与用脑机接口控制一个虚拟3D手臂的表现进行了比较。
“We compared iBCI control of his paralyzed arm to that of a virtual 3D arm”
本研究报告的数据来自第二次FES植入手术之后收集的26次虚拟现实与FES对比测试,以及功能性任务测试。
“Data reported in this study are from 26 VR vs. FES comparison sessions and functional task sessions collected after the second FES implant surgery”
受试者成功地用大脑信号指挥单关节和协调的多关节手臂动作,完成点到点目标获取(准确率80%–100%),先是用虚拟手臂,随后是用FES驱动自己的手臂。
“The participant successfully cortically commanded single-joint and coordinated multi-joint arm movements for point-to-point target acquisitions (80% – 100% accuracy) using first a virtual arm, and second his own arm animated by FES”
受试者用自己瘫痪的手臂,自主地完成了自定节奏的伸手动作,喝到一杯咖啡(在一次测试中12次尝试成功11次),并自己进食。
“Using his paralyzed arm, the participant volitionally performed self-paced reaches to drink a mug of coffee (successfully completing 11 of 12 attempts within a single session) and feed himself”
平均而言,与用大脑控制虚拟手臂相比,受试者控制自己手臂时成功获取目标的次数更少。
“On average, he achieved fewer target acquisitions during control of his own arm as compared to control of the virtual arm (Figure 3B)”
在单关节动作中,约80%的失败试验是由于控制接口的困难,主要是无法保持想要的手部抓握姿势。
“For single-joint movements, ~80% of failed trials were due to control interface challenges, predominantly due to the inability to maintain a desired hand grasp posture (Supplementary Table 2)”
约15%的失败试验是由于肌肉疲劳,而约5%(仅两次试验)是由于未能解码出合适的指挥信号。
“~15% of failed trials were due to muscle fatigue, while ~5% (only two trials) were due to failure to decode an appropriate command signal”
他完成喝咖啡任务需要20–40秒,在所示的那次测试中12次尝试成功了11次。
“He required between 20–40 seconds to complete the drinking task and was successful in 11 of 12 attempts made during the illustrated session (Figure 4B)”
在整个研究过程中,受试者发生了四次轻微的、与设备相关的不良事件(没有严重不良事件),全部得到快速治疗并缓解,并已适当地报告给主管的机构审查委员会(IRB)。
“Over the course of the study, the participant had four minor (and no serious) device-related adverse events, all of which were quickly treated, resolved, and appropriately reported to the governing IRBs”
当FES系统关闭时,受试者完全无法做出有意义的动作;他残存的轻微肩带活动只能引起小而不可控的肘部抽动,完全无法移动手,尽管他试图指挥所需的的手臂动作。
“The participant was completely unable to perform meaningful movements with the FES system turned off (Figure 4C); his minor residual shoulder girdle motion only caused a small, uncontrolled elbow jerk and could not move his hand at all, despite his attempts to command the required arm movements”
采用经皮电极这一选择限制了能够恢复的关节数量和活动范围,而未来完全植入式的FES系统可以利用改进的电极设计和手术放置来弥补。
“While this choice limited the number of joints that could be restored and their ranges-of-motion, future fully implantable FES systems can take advantage of enhanced electrode design and surgical placement”
在脑控制下,FES驱动的动作比虚拟手臂的同样动作要慢一些,也不够准确。
“FES movements were moderately slower and less accurate than the same movements of the VR arm under brain control (Figure 3, Supplementary Figure 4,5)”
虽然脑机接口指挥的系统(包括机械臂)尚未能恢复到与健全人动作相同的速度和精度,但目前能恢复的粗略动作水平仍足以完成有临床意义的功能(如自己进食)。
“While iBCI-commanded systems (including robotic arms) have not yet restored movements with the same speed and precision of able-bodied movements, the current level of gross movement they can restore is still enough to achieve clinically relevant functions (such as self-feeding)”
受试者需要持续用视觉反馈来观察自己的手臂动作,因为缺乏本体感觉使他无法知道手臂的位置,因此在没有视觉反馈时无法完成有意义的动作。
“Finally, note that the participant required continuous visual feedback of his arm movements, as lack of proprioception prevented knowledge of arm position and hence an inability to perform meaningful movements without visual feedback”
受试者(编号T8)加入了BrainGate2先导性临床试验(ClinicalTrials.gov NCT00912041),并对经大学医院克利夫兰医学中心和麻省总医院机构审查委员会批准的研究程序签署了知情同意。
“The participant (ID# T8) enrolled into the BrainGate2 pilot clinical trial (ClinicalTrials.gov NCT00912041) and gave informed consent for study procedures as approved by the Institutional Review Boards of University Hospitals Cleveland Medical Center (Cleveland, OH) and Massachusetts General Hospital (Boston, MA)”
这项研究已在ClinicalTrials.gov注册,注册号为NCT00912041。
“This study is registered with ClinicalTrials.gov, NCT00912041”
这是一项在瘫痪者脑内植入电极、并用电流刺激肌肉的早期(首次人体)概念验证试验,受试者只有一人,属于风险较高、受益尚不确定的研究;审查时要重点权衡他可能获得的益处与手术、植入物和长期刺激带来的风险,并确认研究方案对风险有持续监测和退出安排。AI 分析
原文写明受试者本人签署了知情同意,研究也经过两家医院伦理委员会的批准;建议核实知情同意材料是否用受试者能听懂的语言说明了这是一项探索性研究、效果不保证,以及植入电极日后可能取出或长期留置等关键信息。AI 分析
受试者是颈部脊髓损伤、四肢瘫痪的成年人,研究团队在其瘫痪的手臂上做手术植入电极。需要关注的是:这类受试者因身体依赖他人,同意是否真正自愿;建议核实招募和同意过程中是否有独立于研究团队的人员参与,避免因依赖关系影响决定。AI 分析
原文报告研究期间出现四次轻微的、与装置相关的不良事件,没有严重不良事件,并称都已及时处理、解决并上报伦理委员会。建议核实不良事件的记录是否完整、由谁判定与研究装置相关,以及是否设有独立的数据安全监查机制。AI 分析
这项研究由学术机构主导,涉及多家器械厂商提供植入装置。建议核实研究者与厂商之间的资助、专利或商业利益关系是否已向伦理委员会申报并向受试者披露;同时关注研究成果未来转化为临床产品后,受试者群体能否公平获得。AI 分析
- 四肢瘫痪(tetraplegia)
- 颈段脊髓损伤后,四肢和躯干大部分失去随意运动和感觉功能的状态。
- 功能性电刺激(FES)
- 用电流刺激瘫痪肢体的肌肉或神经,让肌肉收缩产生动作,好比给停摆的机器重新通电。
- 脑机接口(iBCI,皮层内脑机接口)
- 把电极植入大脑运动皮层,直接读取大脑发出“想动”的信号,再由计算机翻译成控制外部设备的指令,相当于在大脑和机器之间架一条直通线。
- ASIA A
- 脊髓损伤严重程度分级中的最高级别损伤,损伤平面以下完全没有运动和感觉功能。
- 经皮电极(percutaneous electrodes)
- 穿过皮肤插入肌肉的细电极,部分留在体外连接刺激器,可以拔除,不同于完全埋在体内的植入式电极。
- 本体感觉(proprioception)
- 不用看也能感知自己肢体位置和运动的感觉,像身体自带的“位置传感器”;失去它就得靠眼睛盯着才能知道手在哪。
这篇还没有动画
动画会把研究的流程、作用机制和关键结果一步一步演示出来,每一步都标明出自原文哪里。制作大约需要一两分钟。
摘要Abstract
Background: People with chronic tetraplegia due to high cervical spinal cord injury (SCI) can regain limb movements through coordinated electrical stimulation of peripheral muscles and nerves, known as Functional Electrical Stimulation (FES). Users typically command FES systems through other preserved, but limited and unrelated, volitional movements (e.g. facial muscle activity, head movements). We demonstrate an individual with traumatic high cervical SCI performing coordinated reaching and grasping movements using his own paralyzed arm and hand, reanimated through FES, and commanded using his own cortical signals through an intracortical brain-computer-interface (iBCI).
Methods: The study participant (53 years old, C4, ASIA A) received two intracortical microelectrode arrays in the hand area of motor cortex, and 36 percutaneous electrodes for electrically stimulating hand, elbow, and shoulder muscles. The participant used a motorized mobile arm support for gravitational assistance and to provide humeral ab/adduction under cortical control. We assessed the participant’s ability to cortically command his paralyzed arm to perform simple single-joint arm/hand movements and functionally meaningful multi-joint movements. We compared iBCI control of his paralyzed arm to that of a virtual 3D arm. This study is registered with ClinicalTrials.gov, NCT00912041.
Findings: The participant successfully cortically commanded single-joint and coordinated multi-joint arm movements for point-to-point target acquisitions (80% – 100% accuracy) using first a virtual arm, and second his own arm animated by FES. Using his paralyzed arm, the participant volitionally performed self-paced reaches to drink a mug of coffee (successfully completing 11 of 12 attempts within a single session) and feed himself.
Interpretation: This is the first demonstration of a combined FES+iBCI neuroprosthesis for both reaching and grasping for people with SCI resulting in chronic tetraplegia, and represents a major advance, with a clear translational path, for clinically viable neuroprostheses for restoring reaching and grasping post-paralysis.
引言INTRODUCTION
High cervical spinal cord injury (SCI) resulting in tetraplegia prevents affected persons from performing reaching and grasping movements required for many activities-of-daily-living. Functional electrical stimulation (FES), in the absence of descending motor commands, applies spatiotemporal patterns of stimulation to peripheral nerves and muscles to reanimate paralyzed limbs for restoring lost functions. FES can be delivered through skin surface, intramuscular, or nerve cuff electrodes, and has successfully restored grasping to persons with mid- to low-level cervical SCI, who retain both volitional shoulder and elbow movements for commanding stimulation.
Restoring multi-joint reaching and grasping is more difficult in persons with high cervical SCI because the few available command options (sip-and-puff, eye tracking, retained head/neck movements) are unintuitive, scale poorly for commanding coordinated multi-joint movements, and interfere with intact head and face function. Intracortical brain-computer-interfaces (iBCIs) that directly map cortical activity to desired movement eschew the need for retained volitional movement, thereby potentially addressing these shortcomings. Intact non-human-primates (NHPs) and humans with paralysis have successfully used iBCIs to command cursor movements and reaching and grasping using robotic limbs. Temporarily paralyzed NHPs have used iBCIs to command implanted FES-actuated wrist and grasping movements. A more recent study used an iBCI coupled with surface electrical stimulation to provide assistive hand grasping to an individual with C5/C6 SCI who retained volitional shoulder and elbow function. However, the 25-year-old Freehand implanted FES system already successfully restored hand grasping to persons who retained volitional arm function, without requiring an iBCI. We for the first time show an individual with chronic tetraplegia using an implanted FES system to make both reaching and grasping movements, intuitively and effectively commanded by an iBCI, with a translational path for future clinical viability.
METHODS / Cortical and FES Implications
The participant (ID# T8) enrolled into the BrainGate2 pilot clinical trial (ClinicalTrials.gov NCT00912041) and gave informed consent for study procedures as approved by the Institutional Review Boards of University Hospitals Cleveland Medical Center (Cleveland, OH) and Massachusetts General Hospital (Boston, MA). At the time of implant, he was a 53-year-old man who experienced traumatic high cervical SCI (C4, ASIA A) eight years before enrollment. On his dominant right side (contralateral to the intracortical implant), he retained limited and nonfunctional voluntary shoulder girdle motion, but no voluntary glenohumeral, elbow, or hand function, and no sensation below the shoulder. An implanted baclofen pump controlled spasticity of his dominant arm.
Two 96-channel microelectrode arrays (Blackrock Microsystems, Salt Lake City, Utah) were implanted into the hand area on the precentral gyrus of his motor cortex (Supplementary Figure 1). After four months of using the iBCI to command movements of a 3D virtual arm, the participant received, during two procedures (125 and 280 days-post-implant), 36 percutaneous muscle stimulating electrodes (Synapse Biomedical, Oberlin, OH) in his right upper and lower arm, including four percutaneous anodic current return electrodes, to restore finger and thumb (for a lateral hand grasp), wrist, elbow, and shoulder movements (Supplemental Table 3 lists implanted muscles). Starting 142 days-post-implant, all implanted muscles were exercised using cyclical electrical stimulation patterns to improve strength, range-of-motion, and fatigue resistance. Exercise occurred 18 out of 45 weeks, averaging 8 hours/week spread over 2–3 days. Over the course of the study, the participant had four minor (and no serious) device-related adverse events, all of which were quickly treated, resolved, and appropriately reported to the governing IRBs.
METHODS / FES+iBCI System Architecture and Neural Decoding
Figure 1A illustrates the FES+iBCI system. The iBCI consists of the implanted recording microelectrode arrays, with a neural decoder that translated recorded cortical activity into command signals for controlling muscle stimulation to produce coordinated reaching and grasping movements. The FES system consists of an external stimulator that delivered charge-balanced, biphasic, constant-current stimulation through percutaneous electrodes to produce muscle contractions and subsequent limb movement. The stimulation had a fixed current amplitude (20 mA) and frequency (12.5 Hz) and a variable pulse duration of 0–200 μs. The current pulse duration (“pulse-width”) applied at a given electrode determined the strength of the muscle contraction. The participant used a Mobile Arm Support (MAS) (Focal MEDITECH, Tilburg, Netherlands) for support against gravity and motorized humeral ab/adduction (also under cortical command), since neither his residual shoulder motion nor deltoid stimulation provided adequate humeral ab/adduction. Instrumented goniometers (Biometrics Ltd.-US, Ladysmith, VA) on his elbow, wrist, and hand measured joint motions.
Neural decoders were calibrated daily at the beginning of each experimental session to translate cortical activity patterns into movement commands for a virtual reality arm or the FES-actuated arm. Daily re-calibration helped to account for day-to-day variability in the recorded activity. The decoders used two neural features from each electrode of the intracortical arrays: 1) unsorted threshold crossing “firing” rates, determined by counting all action potentials in a 20ms time window that crossed a preset noise threshold, and 2) average spectral high frequency power (250–3000 Hz) in a 20ms time window. The decoders used a linear transformation function, similar to the Kalman filter used in recent iBCI applications, to map the neural features to three movement commands. For the virtual arm, decoded commands determined the instantaneous movement velocities for the virtual arm joints (shoulder, elbow, wrist, grasp). For the FES arm, decoded commands determined the change in the percent activation of stimulation patterns associated with elbow, wrist or hand movements (or determined the actuation of the MAS). The stimulation patterns made it easy for the participant to coordinate the activity of multiple electrodes in a graded fashion with only a few command signals (see Supplemental Methods). Figure 1B illustrates example stimulation patterns designed such that increases/decreases in the percent activation of the pattern smoothly coordinated multiple electrodes to cause joint extension/flexion. The decoded command signal for controlling the MAS caused no movement if it was below a certain threshold, and otherwise caused ab/adduction at a constant rate.
METHODS / Session Overview
Figure 2A illustrates the timeline of surgeries and sessions. As part of the BrainGate2 pilot clinical trial, the study participant performed various virtual reality (VR) arm control sessions not directly related to the present study. For this study, he performed two types of sessions: (1) sessions where he controlled both the VR arm and the FES system in order to compare performance between the two, and (2) sessions where he controlled the FES system to complete meaningful functional tasks. Data reported in this study are from 26 VR vs. FES comparison sessions and functional task sessions collected after the second FES implant surgery.
METHODS / Comparing Cortical Control of a Virtual Reality Arm vs FES-Actuated Arm
During the VR vs. FES comparison sessions, the participant cortically commanded single and multi-joint movements of both a 3D virtual arm and his FES arm to perform point-to-point target acquisitions. Since the virtual arm had no inertial or musculoskeletal dynamics, it was useful for demonstrating the quality of the decoded iBCI commands independently of any control difficulties added by the FES-actuated arm and the stimulation patterns.
Experimental sessions consisted of three conditions (Figure 2C). During the attempted movement condition (AM), the participant observed the virtual arm make goal-directed, point-to-point movements while he was instructed to attempt those same movements. Neural activity evoked during these attempted movements was used for decoder initialization. During the virtual reality condition (VR), the participant used this initial decoder to cortically command the real-time velocities of the elbow, wrist, hand, and/or shoulder joints to perform single and multi-joint movements of the opaque virtual arm to a target configuration represented by a translucent arm (Figure 2B). Decoder parameters were updated after each VR block by re-calibrating the decoder with all available data, and then held fixed for the FES condition. During the FES condition, the participant performed single and multi-joint movements of his own FES-actuated arm while receiving visual feedback of the arm movements and the target location via the VR system. The virtual arm’s joint angles were set equal to the joint angles recorded by the instrumented goniometers and the participant looked at the computer monitor during this phase instead of his own FES-actuated arm. This experimental set up (Supplemental Videos 1–2) enabled precise visualization of the target arm configurations during point-to-point FES arm movements and facilitated comparison between VR and FES movements (since visual feedback of the target location and the arm location was identical). The decoder output for the untested joints was set to zero. Supplementary Table 1 summarizes all sessions and joints tested during each day.
METHODS / Functional Task Demonstrations
To demonstrate the potential of the system to restore meaningful function, the participant completed both a coffee drinking and a self-feeding task. For these sessions, we calibrated the decoder while he observed and controlled FES-actuated movements instead of virtual reality movements. First, we initialized the decoder using neural data recorded while his arm was automatically driven by the FES system (i.e. computer-controlled) to make elbow, hand, and shoulder (MAS) movements. He was instructed to simultaneously attempt to make the observed arm motions. We then refined the decoder by using neural data recorded while he performed user-controlled single-joint FES-actuated movements, cued by audio commands instead of the VR game. After refinement, the neural decoder was held constant for the functional tasks. We found that this calibration scheme worked better than calibrating with VR data, potentially because the difference in visual feedback between VR and physical reality caused the neural activity to change.
RESULTS / Quality of Recorded Neural Activity
Neural activity was strongly related to the participant’s intended movement commands during attempted movement (AM), VR, and FES movement conditions. Figure 2D illustrates example neural activity that was strongly tuned to wrist flexion over wrist extension during each condition (substantially more threshold crossings were observed during attempted wrist flexion over wrist extension). Similar consistency of neural tuning between task conditions (AM vs VR vs FES) on some channels was also observed for elbow flexion/extension, humeral ab/adduction, and hand opening/closing (Supplementary Figure 3). Of the 192 electrodes, we identified a neural feature (threshold crossing or spectral high frequency power) that coded for hand opening/closing on 15±2 (mean±SD) electrodes, for elbow flexion/extension on 25±2 electrodes, for wrist flexion/extension on 25±4 electrodes, and for humeral ab/adduction on 27±20 electrodes (Supplemental Figure 8). Supplementary Figure 9 reports the number of isolatable single neurons recorded over time.
RESULTS / Quality of Virtual and FES Arm Movements
Across the VR vs. FES comparison sessions, the participant consistently achieved 80–100% success rates during single-joint movements of the elbow, wrist, hand, and MAS (humeral ab/adduction) to specified target positions (Figure 3, Supplementary Video 1). To successfully acquire a target, he had to maintain the specified joint angle within a certain tolerance around the target for 500ms without exceeding a maximum movement time of 8–12 seconds. The starting position of each movement was equal to the ending position of the previous movement.
For some joint movements (elbow flexion/extension, MAS ab/adduction), the participant acquired targets with his FES-actuated arm as quickly and as successfully as the virtual arm (Figure 3A). For other joint movements (wrist flexion/extension, hand opening/closing), he achieved high success rates but targets were acquired more slowly and speeds varied non-uniformly as a function of joint angle (Figure 3A). Results for two and three-joint movements are shown in Supplementary Figures 4–5 and Supplementary Video 2. On average, he achieved fewer target acquisitions during control of his own arm as compared to control of the virtual arm (Figure 3B). However, FES movements were still far more successful (more target acquisitions) than chance movements for both single and multi-joint movements, and far more successful than what he could achieve with residual voluntary shoulder movements alone (Figure 3B, Supplementary Figure 6, Supplementary Table 4).
Failed reaching attempts were categorized as due to: (1) muscle fatigue making it impossible to reach the target even at full stimulation, (2) control interface (FES and MAS) challenges making it difficult to accurately stop within the target region, or (3) failure to decode the correct command signal to move the joint towards the target. Supplementary Figure 7 illustrates each failure mode. For single-joint movements, ~80% of failed trials were due to control interface challenges, predominantly due to the inability to maintain a desired hand grasp posture (Supplementary Table 2). These trials occurred primarily when the decoded commands for the hand were mapped to highly nonlinear portions of the stimulation pattern. Some portions of the pattern contained “dead space” that did not move the hand very much, while other portions caused large, quick movements, resulting in target overshoots. We alleviated this problem in later blocks by using an automatic procedure that warped the stimulation pattern so that command signals were mapped linearly to equilibrium positions (Supplementary Figure 8). ~15% of failed trials were due to muscle fatigue, while ~5% (only two trials) were due to failure to decode an appropriate command signal.
For multi-joint movements, the dominant failure mode was also control interface challenges (contributed to 67% of failed trials), and was due primarily to MAS movements causing undesired motion of other joints. Interestingly, failure to decode an appropriate command signal for at least one of the joints was more common for multi-joint movements (contributed to 38% of failed trials). Decoding failure may have been more common in multi-joint trials because of the cognitive burden of controlling multiple joints with real dynamics and SCI-musculoskeletal limitations. Muscle fatigue was a reason for failure in ~12% of multi-joint (mostly elbow and wrist) trials.
RESULTS / Performance of Functional Reach-to-Grasp Tasks
The participant was able to successfully acquire a cup of coffee and take a drink (Figure 4A, Supplementary Video 4) and feed himself using the FES+iBCI system (Supplementary Video 5). The coffee drinking task required him to 1) extend his elbow, 2) open his hand, 3) grasp the cup securely, 4) flex his elbow to transport it close to his mouth, 5) take a drink using the straw, 6) extend his elbow to return the cup, and 7) release his grasp. He required between 20–40 seconds to complete the drinking task and was successful in 11 of 12 attempts made during the illustrated session (Figure 4B). During self-feeding, he consistently and repeatedly scooped forkfuls of mashed potatoes, and navigated his hand to his mouth to take several bites. When asked to describe how he commanded the FES arm movements, he replied, “It’s probably a good thing that I’m making it move without having to really concentrate hard at it. … I just think ‘out’ and … it just goes”.
The participant was completely unable to perform meaningful movements with the FES system turned off (Figure 4C); his minor residual shoulder girdle motion only caused a small, uncontrolled elbow jerk and could not move his hand at all, despite his attempts to command the required arm movements. This indicates that no substantial motor recovery occurred due to the FES and/or iBCI (demonstrated more extensively in Supplementary Figure 6). Supplementary video 3 offers an additional qualitative comparison of the movements the participant could make with and without the FES+iBCI system, and also shows that he could move each joint individually with the FES+iBCI system while suppressing the motion of undesired joints. Finally, note that the participant required continuous visual feedback of his arm movements, as lack of proprioception prevented knowledge of arm position and hence an inability to perform meaningful movements without visual feedback.
讨论DISCUSSION
FES+iBCI restoration of arm and hand functions, combined with a mobile arm support (MAS) (standard for persons with C4 SCI implanted with FES arm systems) under iBCI command, represents a neurotechnology-based circumvention of SCI, giving persons with chronic tetraplegia intuitive control over reaching and grasping movements using their paralyzed limbs. The study significantly extends previous iBCI research with individuals with paralysis controlling cursors or robotic limbs. The movements afforded to the study participant (reaching out, grasping/scooping, reaching back to the face) allowed him to repeatedly take a drink of coffee and feed himself with his own arm and hand, solely of his own volition. These actions are representative of movements needed to perform a wide range of reaching tasks, suggesting that more functional activities are achievable with the current system.
FES movements were moderately slower and less accurate than the same movements of the VR arm under brain control (Figure 3, Supplementary Figure 4,5). This discrepancy may have been due to the difference in time spent practicing VR vs FES (65 vs. 15 hours), but may have also been caused by the more difficult control task presented by an FES-activated arm: dynamics due to arm mass, muscle contractile properties, interactions between joints, and MAS motor dynamics. We previously demonstrated that these control difficulties are addressable with a feedback control system that converts higher level movement commands decoded from the iBCI (e.g. desired joint velocities) into the lower level muscle stimulations needed to smoothly achieve that movement. The feedback controller incorporates joint angle sensors that continuously sense any movement “error” or deviation from the desired movement, and recruits the appropriate muscles to reduce that error while taking into account the dynamics of the musculoskeletal limb (similar to how robotic arms are controlled). It should be noted that even without an implemented feedback controller, the participant was able to modulate his neural activity and use visual feedback alone to perform meaningful FES arm movements, even on day one.
The percutaneous, readily removable, FES electrodes provide proof-of-concept for fully-implanted FES systems. While this choice limited the number of joints that could be restored and their ranges-of-motion, future fully implantable FES systems can take advantage of enhanced electrode design and surgical placement (e.g., more precisely located intramuscular electrodes implanted via open surgery, or peripheral nerve cuff electrodes for more distributed motor unit recruitment) and associated techniques (e.g., model-based optimization of muscle stimulation patterns, muscle tendon transfers to replace the functions of denervated muscles, and more extensive exercise programs) to restore motion more fully. The use of implanted FES is critical for clinical adoption of this technology. While some earlier studies have focused on the use of surface FES to restore only hand grasp to persons with lower-level SCI, either commanded by electroencephalography (EEG) or iBCIs, surface FES systems do not have a history of wide spread and long term clinical adoption. In contrast, fully implanted systems for FES grasp restoration, specifically the Freehand, have a history of successful clinical adoption, likely due to the seamlessness of day-to-day setup and use, and their durability (<1% of electrodes fail over 3 years).
While iBCI-commanded systems (including robotic arms) have not yet restored movements with the same speed and precision of able-bodied movements, the current level of gross movement they can restore is still enough to achieve clinically relevant functions (such as self-feeding). Enhanced speed, precision, and multifunctional control may be achievable through electrode technologies that record more neurons from distributed cortical networks, improved decoding algorithms, implantable FES technologies, and restored somatosensation. Somatosensation restored through intracortical stimulation may also eventually allow users to make reach and grasp movements that are safer in the absence of constant visual feedback, due to sensory feedback of object properties (e.g. temperature) in the reachable workspace. Despite current limitations, iBCIs currently offer the best option for seamless clinical integration and greater functional performance, particularly over their non-invasive counterparts (e.g. continuous control of a high-dimensional robotic limb has been successfully demonstrated with iBCI systems but never with EEG). Research advances in intracortical electrode biocompatibility and fully-implanted brain recording interfaces continue to increase the clinical viability of iBCI-commanded systems.
The present FES+iBCI system offers persons with chronic tetraplegia from SCI the possibility of regaining lost arm and hand function to perform activities-of-daily-living. Continued clinical translation of this technology will be aided by iBCI and FES technological advances resulting in smoother and more dexterous arm and hand movements. Future systems inspired by this work may provide full-time and more accurate control of the arm and hand, enabling restoration of a wider range of functional activities and resulting in increased independence and quality-of-life.
Figures and tables
Overview of the FES+iBCI System. (A) Neural activity is recorded from two microelectrode arrays implanted in the motor cortex. The recorded activity is then decoded into command signals that control the stimulation of biceps, triceps, forearm, and hand muscles, as well as the actuation of a mobile arm support (MAS), to enable cortical control of whole arm movements. Muscle stimulation was performed through percutaneous intramuscular fine-wire electrodes, and instrumented goniometers (Biometrics Ltd.-US, Ladysmith, VA) quantified the resultant wrist, elbow, and hand aperture movements, while an orientation sensor quantified MAS movements. (B) Simulation patterns convert the decoded command signals into the appropriate pulse widths to apply to each individual FES electrode, enabling the participant to coordinate the action of multiple electrodes and muscles using only a single command. Example stimulation patterns for the elbow, wrist, and hand are shown; supplementary Figure 8 illustrates how a stimulation pattern controls the angle of a joint.
Session overview. (A) Timeline of implants and experimental sessions. All days are referenced to the day of cortical implant (December 1, 2014 – 0 Days Post Implant). (B) Example image from the virtual reality game. The virtual arm is opaque while the target arm configuration (wrist flexion in this case) is translucent. (C) During the VR vs. FES comparison sessions, the participant completed three different experimental conditions. Block diagrams of each condition and an example session timeline are shown. (D) Example raster plots showing the timing of threshold crossings (top rows) and the average threshold crossing rates (bottom row) of a single channel tuned to wrist flexion and extension during a single-joint wrist movement task. The dotted line at t=0 indicates the presentation of the target movement. This channel records more threshold crossings when flexion targets are presented and has similar tuning properties during all three experimental conditions.
Single-joint FES and mobile arm support (MAS) movements under real-time brain control. (A) (1st column) Restored arm and hand movements and achievable ranges of motion. Line drawings are made from actual photographs of restored movements, and show complete range of restored motion. (2nd–4th columns) Overlaid time series of joint motions towards each target (columns) during an example block of each movement (rows). Each line illustrates a single movement from the example FES block (blue) or virtual reality block (pink). Gray rectangles illustrate the target and the tolerance allowed for target acquisition. Target distances (from the flexion to extension target) and allowed tolerances (widths) were 43.4°±6.0° (elbow), 24°±3.4° (wrist), 35.8°±5.8° (hand), and 41.3°±5.1° (MAS). The participant was in full control of the joint at all times (the joint position was not reset after a target was acquired). Example blocks with high success rates were chosen for illustration. (B) Success rate and average movement time is summarized for each FES block (circles). Circles are different colors if they occurred on different days. Average virtual reality performance (blue dotted line) and chance performance (red dotted line) are shown for reference. Supplementary Table 4 gives a more detailed quantification with accompanying statistical tests.
The participant using the FES+iBCI system to take a drink of coffee. (A) He reached out to grasp the cup of coffee (left) and bringing it to his mouth to take a drink through a straw (right). Photos taken on trial day 392 (2015.12.28). (B) The length of time it took to complete each phase of the drinking task. Data is shown for 12 trials completed within a single experimental session; only one trial was failed when the cup was dropped. (C) Example time series of elbow and hand motion when the FES+iBCI system was turned on (left) and when the FES system was turned off (right). When the system was on, the decoded neural commands (blue) and the elbow and hand joint angles (orange) changed appropriately as the participant moved through the phases of the task, enabling him to take a drink of coffee. When the system was off, he could only make small, uncontrolled elbow jerks caused by his residual shoulder motion and could not move his hand at all. Data for panels B and C collected on trial day 463 (2016.03.08).
还没有查过关联研究
我会去找这篇研究之前的基础工作、做类似事情的研究,以及之后引用它的研究,并说明每篇为什么相关。