自主人工智能驱动的糖尿病视网膜病变即时筛查与读片中心多专家复核对比:AEYE-DS 三项前瞻性对照关键验证研究,逾 1200 例患者
Autonomous AI-driven point-of-care screening for diabetic retinopathy compared to reading center multi-expert clinical review: results from three prospective controlled pivotal validation studies with AEYE-DS in over 1,200 patients
作者:Zack Dvey-Aharon, Moshe Livne, Dan Margalit, Amit Wohl, Adi Haupt, Rachelle Aviv, Ahava Stein, Adi Rachelson, Jonathan Nussdorf, Tsontcho Ianchulev
Front Digit Health · 2026年8月25日 · Dvey-Aharon 等 10 位作者
不需要生物学背景,多打比方
正在获取全文并生成讲解(拿不到全文就依据摘要)…
已等待 0 秒大约需要 10–20 秒
可以先看别的,做好了会自动出现在这里。
这篇还没有动画
动画会把研究的流程、作用机制和关键结果一步一步演示出来,每一步都标明出自原文哪里。制作大约需要一两分钟。
摘要Abstract
PURPOSE: Diabetic retinopathy (DR) remains the leading cause of blindness among working-age adults and requires regular screening to detect progression among the growing global diabetic population. This study evaluated the performance of AEYE-DS, an autonomous artificial intelligence (AI) system designed for high-throughput, point-of-care analysis of retinal images, in detecting more-than-mild diabetic retinopathy (mtmDR) during routine screening of patients with diabetes who had not previously been diagnosed with DR.
PRINCIPAL RESULTS: AEYE-DS was tested across three prospective clinical studies using two FDA-cleared non-mydriatic retinal cameras: the handheld Aurora and the desktop Topcon NW400. The algorithm autonomously analyzed retinal images and determined mtmDR presence. Diagnostic outcomes were compared to a reference standard based on the Early Treatment for Diabetic Retinopathy Study (ETDRS) severity grading performed by multi-expert review at an independent reading center. Sensitivity and specificity were 93% and 91% in AEYE-1 (95% CI: 83%-97% and 88%-94%), 92% and 94% in AEYE-2 (95% CI: 79%-97% and 90%-96%), and 93% and 89% in AEYE-3 (95% CI: 80%-97% and 85%-92%). Imageability was >99% in all studies. Intra-operator repeatability exceeded 99% for both devices. Between-operator reproducibility was 98% for the desktop camera and 95% for the handheld device, while between-device reproducibility reached 99% and 97%, respectively.
CONCLUSIONS: AEYE-DS demonstrated high diagnostic accuracy, imageability, reliability, and reproducibility across different operators and devices in non-mydriatic settings. Findings support autonomous AI system use for scalable, point-of-care DR screening, potentially expanding access, streamlining workflows, and reducing the global burden of diabetic eye disease.