一种新型听觉神经接口的刺激稳定性与临床前台架测试
Stimulation-stability and preclinical bench testing of a new auditory nerve interface
作者:Taylor Stump, Stefan Strahl, Joseph Crew, Moritz Leber, Keno Hübner, Sandeep Negi, Justin Vankirk, Meredith E Adams, Thomas Lenarz, Florian Solzbacher, Hubert H Lim, Loren Rieth
Bioelectron Med · 2026年10月14日 · Stump 等 12 位作者
不需要生物学背景,多打比方
正在获取全文并生成讲解(拿不到全文就依据摘要)…
已等待 0 秒大约需要 10–20 秒
可以先看别的,做好了会自动出现在这里。
这篇还没有动画
动画会把研究的流程、作用机制和关键结果一步一步演示出来,每一步都标明出自原文哪里。制作大约需要一两分钟。
摘要Abstract
OBJECTIVE: Cochlear implants are the flagship sensory neural prosthetic, enabling improved hearing and speech perception in >1 million patients with hearing loss. Despite over three decades of active work and innovations in cochlear implants, hearing performance for complex tasks such as speech-in-noise and music appreciation are still quite limited, putatively due to the challenge of selectively stimulating the auditory nerve with intracochlear electrodes through the bony cochlear wall. Our approach is to more selectively evoke hearing percepts through the development of an Auditory Nerve Implant (ANI) device that uses penetrating microelectrodes placed directly in the nerve for stimulation, and which will be evaluated through psychophysical testing as part of a first-in-human study. Stability and integrity of ANI electrodes during passive soaking and stimulation are needed to support initiation of the clinical trial, which we evaluated through thermal accelerated saline soak testing and stimulation stability testing.
METHODS: The ANI electrode stability in saline was assessed by accelerated aging at ≥ 67 [Formula: see text]C with a 1.5-year equivalent endpoint using electrode impedance and microscopy to characterize outcomes. Stimulation stability testing evaluated electrode integrity after 6.3 billion pulses at 100 μA, which is the maximum clinical stimulation intensity, for a one-year equivalent use, and 1200 μA, for a high-dose protocol. Electrode impedance, longitudinal stimulation voltage, and microscopy outcomes were used to the quantify results.
RESULTS: All electrodes had stable impedances around 2 kΩ at the end of the accelerated soak test, except for three with higher impedances but still below the 70 kΩ failure criteria. All but one electrode survived stimulation stability testing. Optical and electron microscopy largely showed little to no change during testing, indicating no observable tip degradation, with the exception of one soak-test device.
CONCLUSIONS: The outcomes from the studies indicate that ANI electrodes exhibit a high degree of integrity and impedance stability. Microscopy characterization proved effective for evaluating electrodes, providing complementary information to standard Backscattered Scanning Electron Microscopy (BSEM). This study suggests the ANI device tolerate a 1.5 year in-dwelling period for a 1 year perceptual study of stimulation evoked percepts, paving the way for clinical translation.