动态心电监测方式在儿童心律失常检出与管理中的作用

JAMA Pediatr · 2026年10月19日 · Hsu 等 8 位作者

问这篇
一分钟了解要点单中心研究比较三种动态心电方式,7天连续监测检出率明显更高。结果7天连续监测检出重大心律失常的比例为6.2%,明显高于24小时动态心电和事件记录器的2.3%;灵敏度73%对27%,阴性预测值高达97.5%,并带来更多治疗方案的改变。

不需要生物学背景,多打比方

正在获取全文并生成讲解(拿不到全文就依据摘要)…

已等待 0 秒大约需要 10–20 秒

可以先看别的,做好了会自动出现在这里。

这篇还没有动画

动画会把研究的流程、作用机制和关键结果一步一步演示出来,每一步都标明出自原文哪里。制作大约需要一两分钟。

目前只拿到了摘要全文暂时拿不到(可能不是免费全文)。下面是论文摘要。

摘要Abstract

摘要第 1 段问这一段

IMPORTANCE: The diagnostic efficacy of different ambulatory electrocardiographic monitoring modalities in this patient population remains unclear.

摘要第 2 段问这一段

OBJECTIVE: To compare the diagnostic yield and clinical utility of 3 ambulatory electrocardiographic modalities in pediatric patients.

摘要第 3 段问这一段

DESIGN, SETTING, AND PARTICIPANTS: This single-center study took place at an outpatient clinic and an inpatient ward of National Taiwan University Children's Hospital. Patients younger than 18 years who underwent ambulatory electrocardiographic monitoring between December 2022 and January 2025 were included. These data were analyzed February 2025 to March 2026.

摘要第 4 段问这一段

EXPOSURE: A continuous 7-day Holter device that functioned as an event recorder was used. Recordings from the first 24 hours were defined as the 24-hour Holter results.

摘要第 5 段问这一段

MAIN OUTCOMES AND MEASURES: Arrhythmias were categorized as major (tachyarrhythmia or bradyarrhythmia) or minor (extrasystolic arrhythmia). All patients were followed up for 1 year; the final arrhythmia diagnosis was used as the reference standard. The primary and secondary outcomes were detection of arrhythmia and a change in clinical management attributable to monitoring findings, respectively.

摘要第 6 段问这一段

RESULTS: In total, 435 children (235 female [54%] and 200 male [46%]; mean [SD] age, 12.3 [3.9] years) were included. The mean (SD) duration of analyzable ambulatory electrocardiographic data was 4.9 (1.7) days. Of 411 children who activated the event recorder, 10.7% had symptom-correlated arrhythmias. The 7-day Holter monitoring demonstrated a significantly higher rate of major arrhythmias (6.2%) than the 24-hour Holter or 7-day event recordings (2.3%; rate difference 95% CI, 1.9%-6.0%; P < .001). Seven-day Holter monitoring showed superior sensitivity (7-day Holter monitoring: 73%; rate difference 95% CI, 54.1%-84.6%; 24-hour Holter monitoring: 27%; rate difference 95% CI, 13.4%-43.1%; and 7-day event recorder: 27%, rate difference 95% CI, 13.4%-43.1%; P < .001) and a high negative predictive value (97.5%, 95% CI, 95.2%-98.6%). Tachyarrhythmias were associated with higher rates of management changes; management changed in 7.1% of patients with 7-day Holter monitoring, 1.6% with 24-hour Holter (rate difference 95% CI, 3.0%-7.6%; P < .001) and 2.8% with a 7-day event recorder (rate difference 95% CI, 2.0%-6.2%; P < .001).

摘要第 7 段问这一段

CONCLUSIONS AND RELEVANCE: In this study of pediatric patients without structural heart disease who underwent electrocardiographic monitoring, 7-day continuous ambulatory electrocardiographic monitoring was associated with a significantly higher diagnostic yield than 24-hour Holter and 7-day event recording. Continuous monitoring may be preferred to detect tachyarrhythmias likely to influence clinical management.