用于人乳头瘤病毒快速检测的集成微流控 CRISPR 诊断装置
Integrated Microfluidic CRISPR-Based Diagnostic Device for Rapid Testing of HPVs
作者:Tianci Wang, Lingyun Li, Yunxiang Wang, Yongjun Han, Teng Li, Chaoran Yu, Yufeng Lou, Zhen Rong, Kai Lin, Shengqi Wang
Biosensors (Basel) · 2026年10月11日 · Wang 等 10 位作者
不需要生物学背景,多打比方
正在获取全文并生成讲解(拿不到全文就依据摘要)…
已等待 0 秒大约需要 10–20 秒
可以先看别的,做好了会自动出现在这里。
这篇还没有动画
动画会把研究的流程、作用机制和关键结果一步一步演示出来,每一步都标明出自原文哪里。制作大约需要一两分钟。
摘要Abstract
Screening for high-risk human papillomavirus (HR-HPV) is a pivotal strategy for the prevention of invasive cervical cancer, particularly in resource-constrained settings. In this study, we developed a CRISPR-based compact automated real-time detection system, termed Ccard, for rapid testing of HPV16, HPV18, and the internal reference gene GAPDH through a streamlined workflow. A simple room-temperature treatment efficiently releases nucleic acids directly from swab samples, bypassing the need for extraction. The mixture was driven through the channels and loaded into the reaction chambers on the microfluidic chip by a vacuum pump. The lyophilized reagents in the reaction zone of the microfluidic chip were redissolved. The system subsequently performs a single-step RPA-CRISPR/Cas13a reaction, with real-time detection enabled by an Internet of Things device. The assay achieved detection within 30 min, with a limit of detection (LOD) of 5 copies/µL for both HPV16 and HPV18. Validation with clinical samples demonstrated high sensitivity (96.8% for HPV16 and 100% for HPV18) and 100% specificity. These results highlight the reliability, precision, and strong potential of the Ccard system for rapid testing of HR-HPVs.