Continuous Glucose Monitoring- Measured Glucose Levels During Oral Glucose Tolerance Testing in Youths With Obesity: Accuracy and Clinical Implications
作者:Claudia Piona, Eleonora Maria Aiello, Erika Caiazza, Aditya Ravi, Antonella Soglia, Elisa Morotti, Giuseppe Riccardi, Claudio Maffeis
Diabetes Care · 2026年11月16日 · Piona 等 8 位作者
不需要生物学背景,多打比方
正在获取全文并生成讲解(拿不到全文就依据摘要)…
已等待 0 秒大约需要 10–20 秒
可以先看别的,做好了会自动出现在这里。
这篇还没有动画
动画会把研究的流程、作用机制和关键结果一步一步演示出来,每一步都标明出自原文哪里。制作大约需要一两分钟。
摘要Abstract
OBJECTIVE: This study aimed to evaluate the accuracy of continuous glucose monitoring (CGM)-derived glucose values versus venous plasma glucose values during oral glucose tolerance testing (OGTT) in youths with obesity, their effect on glycemic classification, and whether deconvolution-based reconstruction improves agreement.
RESEARCH DESIGN AND METHODS: Twenty-nine youths with obesity underwent OGTT while wearing a FreeStyle Libre 2 sensor. Raw CGM values were temporally aligned with venous plasma glucose values and used to reconstruct plasma profiles via a first-order plasma-interstitial kinetic model and inverse filtering. Accuracy, agreement, consensus error grid, and glycemic classification were compared with collected venous plasma glucose values.
RESULTS: Temporally aligned CGM-derived values showed limited agreement with collected venous plasma glucose values (mean absolute relative difference [MARD] 12.96%; 51.72% of values within ±10 mg/dL; 20.06% of values differing by >25 mg/dL) and caused classification shifts. After reconstruction, MARD decreased to 5.55%, the proportion of values within ±10 mg/dL increased to 75.86%, almost all values fell within consensus error grid zone A, and classification shifts were reduced.
CONCLUSIONS: Deconvolution-based CGM reconstruction may support plasma-referenced, less invasive OGTT-based dysglycemic assessment.